8 Specifications

Capacitance Specifications

Table 7-8Capacitance specifications

 

 

 

Accuracy

Display update rate

Function

Range

Resolution

± (% of reading +

(approx)

 

 

 

Offset Error)

 

 

 

 

 

10.000 nF

0.001 nF

1 % + 8

 

 

 

 

 

 

 

100.00 nF

0.01 nF

 

 

 

 

 

 

 

 

1000.0 nF

0.1 nF

 

 

 

 

 

 

 

Capacitance

10.000 µF

0.001 µF

 

4 times/sec.

100.00 µF

0.01 µF

1 % + 5

 

 

 

 

 

 

 

 

1000.0 µF

0.1 µF

1 time/sec.

 

 

 

 

 

 

 

 

10.000 mF

0.001 mF

 

0.1 times/sec.

 

 

 

 

 

 

100.00 mF

0.01 mF

3 % + 10

0.01 times/sec.

 

 

 

 

 

Temperature Specifications

Table 7-9Temperature specifications

Thermocouple

 

 

 

Accuracy

Range

 

Resolution

± (% of reading + No. of

Type

 

 

 

 

Least Significant Digit)

 

 

 

 

K

–200 – 1372 °C/

0.1 °C/

0.3 % + 3 °C/

–328 – 2502

°F

0.1 °F

0.3 % + 6 °F

 

J [2]

–210 – 1200 °C/

0.1 °C/

0.3 % + 3 °C/

–346 – 2192 °F

0.1 °F

0.3 % + 6 °F

 

 

 

 

 

 

Notes for temperature specifications:

1Accuracy of specifications is subject to the following conditions:

The accuracy does not include the tolerance of the thermocouple probe. The thermal sensor plugged into the meter should be placed in the operating environment for at least an hour.

Use the Null function to reduce the thermal effect. Before using the NULL function, set the meter to no ambient

compensation ( ) mode and keep the thermocouple probe as close to the meter as possible, avoiding contact with any surface that has a different temperature from the ambient temperature.

When measuring temperature with respect to any temperature calibrator, try to set both the calibrator and the meter with an external reference (without internal ambient compensation). If both the calibrator and the meter are set with internal reference (with internal ambient compensation), deviation may show between the readings of the calibrator and the meter, due to differences in ambient compensation between the calibrator and the meter.

2This thermocouple type is only available for the U1252A.

156

Agilent U1251A/U1252A User’s and Service Guide

Page 176
Image 176
Agilent Technologies U1251A manual Capacitance Specifications, Temperature Specifications, 8Capacitance specifications