Performance Tests and Calibration | 7 |
Basic Operating Test
The Basic Operating Test is to test the basic operability of the instrument. Repair is required if the instrument fails the Basic Operating Test.
Backlit test
Press the Bat button to test the backlight. It will momentarily toggle backlit ON and OFF.
Testing the display
Press the Hold button and turn on the meter to view all segments of the display. Compare the display with the example in Figure 6- 1.
Figure 6-1 LCD display