Features and Functions

3

Decibel Display

The dBm operation calculates the power delivered to a reference resistance relative to 1 mW and can be applied to DC V, AC V, and AC + DC V measurements for decibel conversion. The voltage measurement is converted to dBm by using the following formula:

The reference resistance may be selected from 1~9999 Ω in

Setup mode. The default value is 50 Ω.

The decibel of voltage is calculated with respect to 1 V. The equation for the voltage measurement is as follows:

dBV = 20 log10 Vin

1 At the , , or rotary switch position,

press to scroll to the dBm measurement on the primary display. The secondary display indicates the AC voltage measurement.

 

N O T E

If the rotary switch is at the “~ V” position, press

to switch

 

between the dBV and the dBm measurements. The dBm or the dBV

 

 

 

 

measurements can be selected at the ACV position. The selection will be

 

 

 

 

the reference for other voltage measurements.

 

2 Press for more than 1 second to exit this mode.

Agilent U1251A/U1252A User’s and Service Guide

57