7 Performance Tests and Calibration

Step

 

 

Test Function

Range

5520A Output

Error from nominal 1 year

 

 

 

 

 

 

U1251A

U1252A

 

 

 

 

 

 

 

 

 

 

 

 

 

 

19

Press

button to go to

5 A

5 A, 1 kHz

± 42 mA

± 37 mA

 

 

 

A mode

 

 

 

 

 

 

 

 

3A

3 A, 5 kHz

± 96 mA

± 96 mA

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

10 A [11]

10 A, 1 kHz

± 100 mA

± 90 mA

 

 

 

 

Square Wave

Use 53131A

 

 

 

 

 

 

Output

 

 

 

 

 

 

 

 

 

 

20

Turn the rotary switch to the

120 Hz @ 50%

 

N/A

± 26 mHz

 

 

 

position

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4800 Hz @ 50%

 

N/A

± 260 mHz

 

 

 

 

 

 

 

 

 

 

 

Duty Cycle

100 Hz @ 50%

 

N/A

± 0.398% [12]

 

 

 

 

 

 

 

 

 

 

 

 

100 Hz @ 25%

 

N/A

± 0.398% [12]

 

 

 

 

 

 

 

 

 

 

 

 

100 Hz @ 75%

 

N/A

± 0.398% [12]

 

 

 

 

 

 

 

 

 

 

 

 

 

Use 34410A

 

 

 

 

 

 

 

 

 

 

 

 

 

Amplitude

4800 Hz @

 

N/A

± 0.2V

 

 

 

 

99.609%

 

 

 

Notes for verification test:

1The additional error to be added as frequency > 20 kHz and signal input < 10% of range: 300 counts of LSD per kHz.

2The accuracy could be 0.05% + 10, always use relative function to zero thermal effect (short test leads) before measuring the signal.

3The accuracy of 500 W and 5 kW is specified after the Null function.

4For range of 50 MW, the RH is specified for < 60%.

5The accuracy is specified for < 50 nS and after Null function as open test lead.

6All frequency counters are susceptible to error when measuring low voltage, low-frequency signals. Shielding inputs from external noise pickup is critical for minimizing measurement errors.

7Use the Null mode to zero residual.

8The accuracy does not include the tolerance of the thermocouple probe. The thermal sensor plugged into the meter should be placed in the operating enviroment for at least an hour. Use the Null function to reduce the thermal effect.

9Always use relative function to zero the thermal effect with open test lead before measuring the signal. If you do not use Relation function, add 20 digits for accuracy purposes.

1010 A continuous and additional 0.5% to specified accuracy when measuring a signal greater than 10 A~20 A for 30 seconds maximum. After measuring current for > 10 A, to cool down the meter for twice the measuring time you applied before low current measurement.