7 Performance Tests and Calibration
Step |
|
| Test Function | Range | 5520A Output | Error from nominal 1 year | |
|
|
|
| ||||
|
| U1251A | U1252A | ||||
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
19 | Press | button to go to | 5 A | 5 A, 1 kHz | ± 42 mA | ± 37 mA | |
|
|
| A mode |
|
|
|
|
|
|
|
| 3A | 3 A, 5 kHz | ± 96 mA | ± 96 mA |
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
| 10 A [11] | 10 A, 1 kHz | ± 100 mA | ± 90 mA |
|
|
|
| Square Wave | Use 53131A |
|
|
|
|
|
| Output |
|
|
|
|
|
|
|
|
|
| |
20 | Turn the rotary switch to the | 120 Hz @ 50% |
| N/A | ± 26 mHz | ||
|
|
| position |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 4800 Hz @ 50% |
| N/A | ± 260 mHz |
|
|
|
|
|
|
|
|
|
|
| Duty Cycle | 100 Hz @ 50% |
| N/A | ± 0.398% [12] |
|
|
|
|
|
|
|
|
|
|
|
| 100 Hz @ 25% |
| N/A | ± 0.398% [12] |
|
|
|
|
|
|
|
|
|
|
|
| 100 Hz @ 75% |
| N/A | ± 0.398% [12] |
|
|
|
|
|
|
|
|
|
|
|
|
| Use 34410A |
|
|
|
|
|
|
|
|
|
|
|
|
| Amplitude | 4800 Hz @ |
| N/A | ± 0.2V |
|
|
|
| 99.609% |
|
|
|
Notes for verification test:
1The additional error to be added as frequency > 20 kHz and signal input < 10% of range: 300 counts of LSD per kHz.
2The accuracy could be 0.05% + 10, always use relative function to zero thermal effect (short test leads) before measuring the signal.
3The accuracy of 500 W and 5 kW is specified after the Null function.
4For range of 50 MW, the RH is specified for < 60%.
5The accuracy is specified for < 50 nS and after Null function as open test lead.
6All frequency counters are susceptible to error when measuring low voltage,
7Use the Null mode to zero residual.
8The accuracy does not include the tolerance of the thermocouple probe. The thermal sensor plugged into the meter should be placed in the operating enviroment for at least an hour. Use the Null function to reduce the thermal effect.
9Always use relative function to zero the thermal effect with open test lead before measuring the signal. If you do not use Relation function, add 20 digits for accuracy purposes.
1010 A continuous and additional 0.5% to specified accuracy when measuring a signal greater than 10 A~20 A for 30 seconds maximum. After measuring current for > 10 A, to cool down the meter for twice the measuring time you applied before low current measurement.
126 | Agilent U1251A/U1252A User’s and Service Guide |