5 Maintenance
Troubleshooting
WA R N I N G
To avoid electrical shock, do not perform any service unless you are qualified to do so.
| If the instrument fails to operate, check the battery and the |
| test leads. Replace them if necessary. And if the instrument |
| still does not function, check the operating procedures in |
| this manual. When servicing, use only specified replacement |
| parts. Table 5- 5 below will assist you to identify some basic |
| problems and their solutions. |
Table | |
|
|
Malfunction | Troubleshooting procedure |
No LCD display after power ON | • Check the battery. Charge or replace it. |
|
|
No beeper tone | • Check the setup mode to verify if the beeper is set to OFF. Then select the |
| desired driving frequency. |
|
|
Fail on current measurement | • Check the fuse. |
|
|
No charging indication [1] | • Check 440 mA fuse. |
| • Check the external adaptor to verify if the output is DC 24 V and the adaptor is |
| plugged into the charging terminals completely. |
| • Line power voltage (100V~240 VAC 50 Hz/ 60 Hz). |
|
|
Battery life very short after full | • Check if the correct rechargeable battery is used. |
charge/Battery not able to charge | • Try to charge and discharge for 2 or 3 cycles in order to maintain the battery’s |
after | highest capacity. |
| • NOTE: The performance of the rechargeable battery may degrade with time. |
|
|
Fail on Remote control | • Check that the optical side of the cable is connected to the meter and the text |
| side of the cover is up. |
| • Check the baud rate, parity, Data bit, Stop bit (default is 9600, n, 8, 1) |
| • Driver install for |
|
|
Notes for basic troubleshooting procedures table:
1Never turn the rotary switch of the multimeter from the OFF position when it is charging.
108 | Agilent U1251A/U1252A User’s and Service Guide |