7 Performance Tests and Calibration

Recommended Test Equipment

The test equipment recommended for the performance verification and adjustment procedures is listed below. If the exact instrument is not available, substitute calibration standards of equivalent accuracy.

A suggested alternative method would be to use the Agilent 3458A 8½ – Digit Digital Multimeter to measure less accurate yet stable sources. The output value measured from the source can be entered into the instrument as the target calibration value.

Table 6-1Recommended Test Equipment

Application

Recommended equipment

Recommended accuracy requirements

 

 

 

DC Voltage

Fluke 5520A

<1/5 instrument 1 year spec

 

 

 

DC Current

Fluke 5520A

<1/5 instrument 1 year spec

 

 

 

Resistance

Fluke 5520A

<1/5 instrument 1 year spec

 

 

 

AC Voltage

Fluke 5520A

<1/5 instrument 1 year spec

 

 

 

AC Current

Fluke 5520A

<1/5 instrument 1 year spec

 

 

 

Frequency

Agilent 33250A

<1/5 instrument 1 year spec

 

 

 

Capacitance

Fluke 5520A

<1/5 instrument 1 year spec

 

 

 

Duty Cycle

Fluke 5520A

<1/5 instrument 1 year spec

 

 

 

Nanosiemens

Fluke 5520A

<1/5 instrument 1 year spec

 

 

 

Diode

Fluke 5520A

<1/5 instrument 1 year spec

 

 

 

Frequency

Agilent 33250A

<1/5 instrument 1 year spec

Counter

 

 

 

 

 

Temperature

Fluke 5520A

<1/5 instrument 1 year spec

 

 

 

Square Wave

Agilent 53131A and Agilent 34401A

<1/5 instrument 1 year spec

 

 

 

Short

Shorting Plug - Dual banana plug with copper wire

short between 2 terminal

 

 

 

 

 

114

Agilent U1251A/U1252A User’s and Service Guide

Page 134
Image 134