Specifications 8
For current measurements
Table
Input Range | Minimum Sensitivity (R.M.S. sine wave) | |
20 Hz – 20 kHz | ||
| ||
500.00 µA | 100 µA | |
|
| |
5000.0 µA | 250 µA | |
|
| |
50.000 mA | 10 mA | |
|
| |
440.00 mA | 25 mA | |
|
| |
5.0000 A | 1 A | |
|
|
Peak Hold Specifications
Table
Signal Width | Accuracy For DC mV/voltage/current |
Single event > 1 ms | 2% + 400 for all ranges |
|
|
Repetitive > 250 µs | 2% + 1000 for all ranges |
|
|
Agilent U1251A/U1252A User’s and Service Guide | 159 |