Performance Tests and Calibration

7

Table 6-4Adjustment table

Step

 

Test Function

Cal Range

Input

 

Cal Item

 

 

 

 

 

U1251A

 

U1252A

 

 

 

 

 

 

 

 

 

 

 

 

 

1

Turn the rotary switch to the

5 V

0.3 V,1 kHz

0.3000 V

 

0.3000 V

 

 

V position

 

 

 

 

 

 

 

 

 

3 V, 1 kHz

3.0000 V

 

3.0000 V

 

 

 

 

3 V, 50 kHz

N/A

 

3.0000 V

 

 

 

 

 

 

 

 

 

 

 

50 V

3 V, 1 kHz

03.000 V

 

03.000 V

 

 

 

 

30 V, 1 kHz

30.000 V

 

30.000 V

 

 

 

 

30 V, 50 kHz

N/A

 

30.000 V

 

 

 

 

 

 

 

 

 

 

 

500 V

30 V,1 kHz

030.00 V

 

030.00 V

 

 

 

 

300 V,1 kHz

300.00 V

 

300.00 V

 

 

 

 

300 V, 50 kHz

N/A

 

300.00 V

 

 

 

 

 

 

 

 

 

 

 

1000 V

30 V, 1 kHz

0030.0 V

 

0030.0 V

 

 

 

 

300 V, 1 kHz

0300.0 V

 

0300.0 V

 

 

 

 

300 V, 50 kHz

N/A

 

0300.0 V

 

 

 

 

 

 

 

2

Turn the rotary switch to

Short

Dual Banana Plug

SHort

 

SHort

 

V position (for model U1252A),

 

with copper wire

 

 

 

 

 

short between 2

 

 

 

 

to

V position (for model

 

 

 

 

 

 

terminals

 

 

 

 

U1251A)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5 V

3 V

3.0000 V

 

3.0000 V

 

 

 

 

 

 

 

 

 

 

 

50 V

30 V

30.000 V

 

30.000 V

 

 

 

 

 

 

 

 

 

 

 

500 V

300 V

300.00 V

 

300.00 V

 

 

 

 

 

 

 

 

 

 

 

1000 V

1000 V

1000.0 V

 

1000.0 V

 

 

 

 

 

 

 

 

Agilent U1251A/U1252A User’s and Service Guide

137

Page 157
Image 157
Agilent Technologies manual 4Adjustment table, Step Test Function Cal Range Input Cal Item U1251A U1252A