1

The keypad at a glance

The operation of each key is shown below. Pressing a key illuminates a related symbol on the display and emits a beep from the multimeter. Turning the rotary switch to another position resets the current operation of the key.

Figure 1- 7 below shows the keypad of the U1252A. The ms% (Pulse width/Duty cycle), Hz , and frequency counter functions are only available on the U1252A.

3

 

 

 

 

 

 

 

 

 

4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5

2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

6

 

 

 

 

 

 

 

 

 

 

 

1

 

 

 

 

 

7

 

 

 

 

 

Figure 1-7U1252A keypad

Table 1-2Keypad desciption and functions

Button

Function when pressed for less than 1 second

Function when pressed for more than 1 second

 

 

 

 

1

 

acts as a toggle switch to turn backlit ON/OFF.

displays the battery capacity for 3 seconds.

 

 

Backlit automatically turns off after 30s (default) [1].

 

2

 

freezes the measured value. In Data Hold

enters Dynamic Recording mode. Press

 

 

mode, press again to trigger hold of the next measured

again to scroll through Max, Min, Avg, and present

 

 

value. In Refresh Hold mode, reading updates

readings (indicated by MAXMINAVG on display).

 

 

automatically once reading is stable and count setting

 

 

 

is exceeded [1].

 

3

 

saves displayed value as a reference to be

enters 1 ms Peak Hold mode. Press

 

 

subtracted from subsequent measurements. Press

to scroll through Max and Min peak readings.

 

 

again to see the relative value that has been saved.

 

 

 

 

 

 

 

Agilent U1251A/U1252A User’s and Service Guide