To unsecure the instrument without the security code
To unsecure the instrument without the correct security code, follow the steps below.
N O T E | If you do not have a record of the security code, you can try 1234 (the fac- | |||
| tory default code) using the front panel. | |||
| ||||
|
|
| ||
| 1 | Record the last 4 digits of the multimeter’s serial number. | ||
| 2 | Turn the rotary switch to | . | |
| 3 | Press | and | simultaneously to enter the |
|
| Calibration Security Code entry mode. | ||
| 4 | The primary display shows 5555 and the secondary | ||
|
| display shows SECUr. |
| |
| 5 | Press | for more than 1 second to enter the Set | |
|
| Default Security Code mode. The secondary display shows | ||
|
| SEr.no and the primary display shows 5555. |
6Use the editing keys character in the code.
and
to step each
Use and
to select each character.
7Set the code, same as the last 4 digits of the instrument’s serial number.
8Press (Save) to confirm the entry.
9If the 4 digits entered are correct, the secondary display will show PASS.