7 Performance Tests and Calibration

To unsecure the instrument without the security code

To unsecure the instrument without the correct security code, follow the steps below.

N O T E

If you do not have a record of the security code, you can try 1234 (the fac-

 

tory default code) using the front panel.

 

 

 

 

 

1

Record the last 4 digits of the multimeter’s serial number.

 

2

Turn the rotary switch to

.

 

3

Press

and

simultaneously to enter the

 

 

Calibration Security Code entry mode.

 

4

The primary display shows 5555 and the secondary

 

 

display shows SECUr.

 

 

5

Press

for more than 1 second to enter the Set

 

 

Default Security Code mode. The secondary display shows

 

 

SEr.no and the primary display shows 5555.

6Use the editing keys character in the code.

and

to step each

Use and

to select each character.

7Set the code, same as the last 4 digits of the instrument’s serial number.

8Press (Save) to confirm the entry.

9If the 4 digits entered are correct, the secondary display will show PASS.

130

Agilent U1251A/U1252A User’s and Service Guide

Page 150
Image 150
Agilent Technologies U1251A manual Tory default code using the front panel