8 Specifications
AC specifications for U1252A
Table
|
|
|
|
| Frequency |
|
|
|
|
|
| 20 Hz – | 45 Hz – | 1 kHz – | 10 kHz – | 20 kHz – | |
Function | Range | Resolution | 45 Hz | 1 kHz | 10 kHz | 20 kHz | 100 kHz [1] | |
| 50.000 mV | 0.001 mV | 1.5 + 60 | 0.4 + 40 | 0.7 + 40 | 0.75 | + 40 | 3.5 + 120 |
|
|
|
|
|
|
|
|
|
| 500.00 mV | 0.01 mV | 1.5 + 60 | 0.4 + 25 | 0.4 + 25 | 0.75 | + 40 | 3.5 +120 |
|
|
|
|
|
|
|
|
|
| 1000.0 mV | 0.1 mV | 1.5 + 60 | 0.4 + 25 | 0.4 + 25 | 0.75 | + 40 | 3.5 + 120 |
Voltage [2] |
|
|
|
|
|
|
|
|
5.0000 V | 0.0001 V | 1.5 + 60 | 0.4 + 25 | 0.4 + 25 | 0.75 | + 40 | 3.5 + 120 | |
| 50.000 V | 0.001 V | 1.5 + 60 | 0.4 + 25 | 0.4 + 25 | 0.75 | + 40 | 3.5 + 120 |
|
|
|
|
|
|
|
| |
| 500.00 V | 0.01 V | 1.5 + 60 | 0.4 + 25 | 0.4 + 25 | 1.5 + 40 | 3.5 + 120 [3] | |
| 1000.0 V | 0.1 V | 1.5 + 60 | 0.4 + 40 | 0.4 + 40 | 1.5 + 40 [3] | - |
Notes for U1252A AC voltage specifications:
1The additional error to be added as frequency > 20 kHz and signal input <10 %of range: 3 counts of LSD per kHz.
2Input impedance: > 1GHz for 50 mV to 1000 mV. Input impedance is 1.1 MΩ (nominal) with <100 pF for 5 V to 1000 V ranges.
3The input signal is lower than the product of 20,000,000
Table
|
|
|
|
| Frequency |
| |
|
|
| 20 Hz – 45 Hz | 45 Hz – 1 kHz |
| 1 kHz – 20 kHz | 20 kHz – |
Function | Range | Resolution |
| 100 kHz [1] | |||
|
|
|
| ||||
| 500.00 µA [2] | 0.01 µA | 1.0 + 20 | 0.7 + 20 | 0.75 + 20 | 5 + 80 | |
| 5000.0 µA | 0.1 µA | 1.0 + 20 | 0.7 + 20 | 0.75 + 20 | 5 + 80 | |
|
|
|
|
|
|
|
|
Current | 50.000 mA | 0.001 mA | 1.0 + 20 | 0.7 + 20 | 0.75 + 20 | 5 + 80 | |
|
|
|
|
|
|
|
|
| 440.00 mA | 0.01 mA | 1.0 + 20 | 0.7 + 20 | 1.5 + 20 | 5 + 80 | |
|
|
|
|
|
|
|
|
| 5.0000 A | 0.0001 A | 1.5 + 20 [4] | 0.7 + 20 | 3 + 60 | - | |
| 10.000 A [3] | 0.001 A | 1.5 + 20 [4] | 0.7 + 20 |
| < 3 A/5 kHz | |
|
|
|
Notes for U1252B AC current specifications:
1The additional error to be added as frequency > 20 kHz and signal input< 10 % of range: 3 counts of LSD per kHz.
2Input current > 35 µArms.
3Current can be measured from 2.5 A up to 10 A continuously. Add 0.5% to the specified accuracy if the signal measured is in the range of 10 A to 20 A and for a period of up to 30 seconds. After measuring current > 10 A, leave the meter to cool down for a period that is twice the measuring time used before application of low current measurement.
4Input current < 3 Arms.
154 | Agilent U1251A/U1252A User’s and Service Guide |