Specifications 8

Input impedance

 

 

Table 7-19Input Impedance

 

 

 

 

 

Function

Range

Input Impedance

 

 

 

 

50.000 mV

10.00 MΩ

 

 

 

 

500.00 mV

10.00 MΩ

 

 

 

 

1000.0 mV

10.00 MΩ

DC Voltage [1]

 

 

5.0000 V

11.10 MΩ

 

50.000 V

10.10 MΩ

 

 

 

 

500.00 V

10.01 MΩ

 

 

 

 

1000.0 V

10.001 MΩ

 

 

 

 

50.000 mV

10.00 MΩ

 

 

 

 

500.00 mV

10.00 MΩ

 

 

 

 

1000.0 mV

10.00 MΩ

AC Voltage [2]

 

 

5.0000 V

10.00 MΩ

 

50.000 V

10.00 MΩ

 

 

 

 

500.00 V

10.00 MΩ

 

 

 

 

1000.0 V

10.00 MΩ

 

 

 

 

50.000 mV

10.00 MΩ

 

 

 

 

500.00 mV

10.00 MΩ

 

 

 

 

1000.0 mV

10.00 MΩ

AC + DC Voltage [2]

 

 

5.0000 V

11.10 MΩ // 10 MΩ

 

50.000 V

10.10 MΩ // 10 MΩ

 

 

 

 

500.00 V

10.01 MΩ // 10 MΩ

 

 

 

 

1000.0 V

10.001 MΩ // 10 MΩ

Notes for input impedance:

1For 5 V to 1000 V range, the specified input impedance in parallel with 10 MΩ at dual display.

2The specified input impedance (nominal) in parallel with <100 pF.

Agilent U1251A/U1252A User’s and Service Guide

163

Page 183
Image 183
Agilent Technologies U1251A manual Input impedance, 19Input Impedance