Agilent Technologies U1251A manual Adjustments Consideration, Problem and repeat the procedure

Models: U1251A

1 185
Download 185 pages 44.17 Kb
Page 153
Image 153

Performance Tests and Calibration

7

Adjustments Consideration

 

You will need a test input cable, a connectors set, and a

 

Shorting Plug to adjust the instrument.

 

After each adjustment, the secondary display briefly shows PASS. If the

N O T E

calibration fails, the handheld multimeter sounds a beep, and an error

 

 

number is shown in the secondary display. Calibration error messages are

 

described on page 143. In the event of a calibration failure, correct the

 

problem and repeat the procedure.

 

 

 

Adjustments for each function should be performed only in

 

the order shown below.

 

1

Allow the instrument to warm up and stabilize for 5

 

 

minutes before performing the adjustments.

 

2

Ensure that during the adjustments, the low battery

 

 

indicator does not appear. Replace the batteries as soon

 

 

as possible to avoid false reading.

 

3

Consider the thermal effects as you are connecting test

 

 

leads to the calibrator and handheld multimeter. It is

 

 

recommended to wait for one minute before you begin the

 

 

calibration after connecting the test leads.

 

4

During ambient temperature adjustment, be sure to turn

 

 

on the instrument for at least 1 hour with the K- type

 

 

thermocouple connected between the instrument and the

 

 

calibration source.

 

Never turn off the instrument during an adjustment. This may delete the

C A U T I O N

 

calibration memory for the present function.

 

 

 

 

Agilent U1251A/U1252A User’s and Service Guide

133

Page 153
Image 153
Agilent Technologies U1251A manual Adjustments Consideration, Problem and repeat the procedure