Performance Tests and Calibration

7

Step

 

Test Function

Cal Range

Input

 

Cal Item

 

 

 

 

 

U1251A

 

U1252A

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5

Press

button to go to the

50 mV

3 mV, 1 kHz

03.000 mV

 

03.000 mV

 

 

mV mode

 

 

 

 

 

 

 

 

 

30 mV, 1 kHz

30.000 mV

 

30.000 mV

 

 

 

 

30 mV, 50 kHz

N/A

 

30.000 mV

 

 

 

 

 

 

 

 

 

 

 

500 mV

30 mV, 1 kHz

030.00 mV

 

030.00 mV

 

 

 

 

300 mV, 1 kHz

300.00 mV

 

300.00 mV

 

 

 

 

300 mV, 50 kHz

N/A

 

300.00 mV

 

 

 

 

 

 

 

 

 

 

 

1000 mV

30 mV, 1 kHz

0030.0 mV

 

0030.0 mV

 

 

 

 

1000 mV, 1 kHz

1000.0 mV

 

1000.0 mV

 

 

 

 

1000 mV, 50 kHz

N/A

 

1000.0 mV

 

 

 

 

 

 

 

6

Turn the rotary switch to the W

Short

Dual Banana

SHort

 

SHort

 

position

 

Plug with copper

 

 

 

 

 

 

 

wire short

 

 

 

 

 

 

 

between 2

 

 

 

 

 

 

 

terminals

 

 

 

 

 

 

 

 

 

 

 

 

 

 

50 MW

Input terminal

oPEn

 

oPEn

 

 

 

 

open (Remove

 

 

 

 

 

 

 

any test leads

 

 

 

 

 

 

 

and Short Plugs

 

 

 

 

 

 

 

from the input

 

 

 

 

 

 

 

terminal)

 

 

 

 

 

 

 

10 MW

10.000 MW

 

10.000 MW

 

 

 

 

 

 

 

 

 

 

 

5 MW

3 MW

3.0000 MW

 

3.0000 MW

 

 

 

 

 

 

 

 

 

 

 

500 kW

300 kW

300.00 kW

 

300.00 kW

 

 

 

 

 

 

 

 

 

 

 

50 kW

30 kW

30.000 kW

 

30.000 kW

 

 

 

 

 

 

 

 

 

 

 

5 kW

3k W

3.0000 kW

 

3.0000 kW

 

 

 

 

 

 

 

 

 

 

 

500 W

300 W

300.00 W

 

300.00 W

 

 

 

 

 

 

 

 

Agilent U1251A/U1252A User’s and Service Guide

139

Page 159
Image 159
Agilent Technologies manual Agilent U1251A/U1252A User’s and Service Guide 139