Performance Tests and Calibration | 7 |
Step |
| Test Function | Cal Range | Input |
| Cal Item | |
|
|
|
| ||||
| U1251A |
| U1252A | ||||
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
5 | Press | button to go to the | 50 mV | 3 mV, 1 kHz | 03.000 mV |
| 03.000 mV |
|
| mV mode |
|
|
|
|
|
|
|
|
| 30 mV, 1 kHz | 30.000 mV |
| 30.000 mV |
|
|
|
| 30 mV, 50 kHz | N/A |
| 30.000 mV |
|
|
|
|
|
|
|
|
|
|
| 500 mV | 30 mV, 1 kHz | 030.00 mV |
| 030.00 mV |
|
|
|
| 300 mV, 1 kHz | 300.00 mV |
| 300.00 mV |
|
|
|
| 300 mV, 50 kHz | N/A |
| 300.00 mV |
|
|
|
|
|
|
|
|
|
|
| 1000 mV | 30 mV, 1 kHz | 0030.0 mV |
| 0030.0 mV |
|
|
|
| 1000 mV, 1 kHz | 1000.0 mV |
| 1000.0 mV |
|
|
|
| 1000 mV, 50 kHz | N/A |
| 1000.0 mV |
|
|
|
|
|
|
| |
6 | Turn the rotary switch to the W | Short | Dual Banana | SHort |
| SHort | |
| position |
| Plug with copper |
|
|
| |
|
|
|
| wire short |
|
|
|
|
|
|
| between 2 |
|
|
|
|
|
|
| terminals |
|
|
|
|
|
|
|
|
|
|
|
|
|
| 50 MW | Input terminal | oPEn |
| oPEn |
|
|
|
| open (Remove |
|
|
|
|
|
|
| any test leads |
|
|
|
|
|
|
| and Short Plugs |
|
|
|
|
|
|
| from the input |
|
|
|
|
|
|
| terminal) |
|
|
|
|
|
|
| 10 MW | 10.000 MW |
| 10.000 MW |
|
|
|
|
|
|
|
|
|
|
| 5 MW | 3 MW | 3.0000 MW |
| 3.0000 MW |
|
|
|
|
|
|
|
|
|
|
| 500 kW | 300 kW | 300.00 kW |
| 300.00 kW |
|
|
|
|
|
|
|
|
|
|
| 50 kW | 30 kW | 30.000 kW |
| 30.000 kW |
|
|
|
|
|
|
|
|
|
|
| 5 kW | 3k W | 3.0000 kW |
| 3.0000 kW |
|
|
|
|
|
|
|
|
|
|
| 500 W | 300 W | 300.00 W |
| 300.00 W |
|
|
|
|
|
|
|
|
Agilent U1251A/U1252A User’s and Service Guide | 139 |