8 Specifications

Frequency Counter Specifications for U1252A

Table 7-15Frequency counter (divide by 1) specifications

 

 

Accuracy

 

 

Range

Resolution

± (% of reading + No.

Sensitivity

Min. Input Frequency

of Least Significant

 

 

 

 

 

 

Digit)

 

 

99.999 Hz

0.001 Hz

0.02% + 3 [1]

 

 

999.99 Hz

0.01 Hz

 

100 mV R.M.S.

 

9.9999 kHz

0.0001 kHz

 

 

0.5 Hz

 

 

 

 

99.999 kHz

0.001 kHz

0.002% + 5

 

 

 

 

 

 

 

999.99 kHz

0.01 kHz

200 mV R.M.S.

 

< 985 kHz

 

 

 

 

 

9.9999 MHz

0.0001 MHz

 

 

 

 

 

 

 

 

 

 

Table 7-16Frequency counter (divide by 100 [4]) specifications

 

 

Accuracy

 

 

Range

Resolution

± (% of reading + No.

Sensitivity

Min. Input Frequency

of Least Significant

 

 

 

 

 

 

Digit)

 

 

9.9999 MHz

0.0001 MHz

0.002 % + 5,

400 mV R.M.S.

1 MHz

99.99 MHz

0.001 MHz

< 20 MHz

600 mV R.M.S.

 

Notes for frequency counter specifications:

1All frequency counters are susceptible to error when measuring low-voltage, low-frequency signals. Shielding inputs from picking up external noise is critical for minimizing measurement errors. For non-square wave signals, an additional 5 counts need to be added.

2The maximum measurement level is < 30 Vpp.

3The minimum measurement frequency of low frequency is set by power-on option to speed up the measurement rate.

4Shown on the secondary display