7 Performance Tests and Calibration
Using the front panel for adjustments
This section describes the procedures to perform adjustments from the front panel.
Selecting the Adjustment Mode
To unsecure the instrument, see “Unsecuring the instrument for calibration” on page 128 or “To unsecure the instrument without the security code” on page 130. Once unsecured, the reference value will be indicated on the primary display.
Entering Adjustment Values
Use the following adjustment procedure to enter an input calibration value from the front panel:
1Use the edit keys and to select each digit in the primary display.
2Use the and the arrow keys to advance through the digits 0 through 9.
3Press when done to start calibration.