Performance Tests and Calibration

7

Step

 

 

Test Function

Range

5520A Output

Error from nominal 1 year

 

 

 

 

 

 

U1251A

U1252A

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

6

Turn the rotary switch to the

50 mV

50 mV

± 75 μV [2]

± 75 μV [2]

 

 

 

mV position

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

500 mV

500 mV

± 0.2 mV

± 0.175 mV

 

 

 

 

 

 

– 500 mV

± 0.2 mV

± 0.175 mV

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1000 mV

1000 mV

± 0.8 mV

± 0.75 mV

 

 

 

 

 

 

– 1000 mV

± 0.8 mV

± 0.75 mV

 

 

 

 

 

 

 

 

 

7

Press

 

button to go to

50 mV

50 mV, 1 kHz

± 0.34 mV

± 0.24 mV

 

 

 

mV mode [1]

 

 

 

 

 

the

 

 

 

 

 

 

 

 

 

 

 

 

50 mV, 10 kHz

± 0.54 mV

± 0.39 mV

 

 

 

 

 

 

50 mV, 20 kHz

N/A

± 0.415 mV

 

 

 

 

 

 

50 mV, 30 kHz

± 0.86 mV

N/A

 

 

 

 

 

 

45 mV, 100 kHz

N/A

± 1.695 mV

 

 

 

 

 

 

 

 

 

 

 

 

 

 

500 mV

500 mV, 45 Hz

± 3.25 mV

± 2.25 mV

 

 

 

 

 

 

500 mV, 1 kHz

± 3.25 mV

± 2.25 mV

 

 

 

 

 

 

500 mV, 10 kHz

± 5.4 mV

± 2.25 mV

 

 

 

 

 

 

500 mV, 20 kHz

N/A

± 4.15 mV

 

 

 

 

 

 

500 mV, 30 kHz

± 8.6 mV

N/A

 

 

 

 

 

 

450 mV, 100 kHz

N/A

± 16.95 mV

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1000 mV

1000 mV, 1 kHz

± 8.5 mV

± 6.5 mV

 

 

 

 

 

 

1000 mV, 10 kHz

± 12.5 mV

± 6.5 mV

 

 

 

 

 

 

1000 mV, 20 kHz

N/A

± 11.5 mV

 

 

 

 

 

 

1000 mV, 30 kHz

± 20.0 mV

N/A

 

 

 

 

 

 

1000 mV, 100 kHz

N/A

± 47 mV

 

 

 

 

 

 

 

 

 

Agilent U1251A/U1252A User’s and Service Guide

123

Page 143
Image 143