Performance Tests and Calibration | 7 |
Step |
|
| Test Function | Range | 5520A Output | Error from nominal 1 year | ||
|
|
|
| |||||
|
| U1251A | U1252A | |||||
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
| |
6 | Turn the rotary switch to the | 50 mV | 50 mV | ± 75 μV [2] | ± 75 μV [2] | |||
|
|
| mV position |
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
| 500 mV | 500 mV | ± 0.2 mV | ± 0.175 mV |
|
|
|
|
|
| – 500 mV | ± 0.2 mV | ± 0.175 mV |
|
|
|
|
|
|
|
|
|
|
|
|
|
| 1000 mV | 1000 mV | ± 0.8 mV | ± 0.75 mV |
|
|
|
|
|
| – 1000 mV | ± 0.8 mV | ± 0.75 mV |
|
|
|
|
|
|
|
|
|
7 | Press |
| button to go to | 50 mV | 50 mV, 1 kHz | ± 0.34 mV | ± 0.24 mV | |
|
|
| mV mode [1] |
|
|
|
| |
| the |
|
|
|
|
|
| |
|
|
|
|
|
| 50 mV, 10 kHz | ± 0.54 mV | ± 0.39 mV |
|
|
|
|
|
| 50 mV, 20 kHz | N/A | ± 0.415 mV |
|
|
|
|
|
| 50 mV, 30 kHz | ± 0.86 mV | N/A |
|
|
|
|
|
| 45 mV, 100 kHz | N/A | ± 1.695 mV |
|
|
|
|
|
|
|
|
|
|
|
|
|
| 500 mV | 500 mV, 45 Hz | ± 3.25 mV | ± 2.25 mV |
|
|
|
|
|
| 500 mV, 1 kHz | ± 3.25 mV | ± 2.25 mV |
|
|
|
|
|
| 500 mV, 10 kHz | ± 5.4 mV | ± 2.25 mV |
|
|
|
|
|
| 500 mV, 20 kHz | N/A | ± 4.15 mV |
|
|
|
|
|
| 500 mV, 30 kHz | ± 8.6 mV | N/A |
|
|
|
|
|
| 450 mV, 100 kHz | N/A | ± 16.95 mV |
|
|
|
|
|
|
|
|
|
|
|
|
|
| 1000 mV | 1000 mV, 1 kHz | ± 8.5 mV | ± 6.5 mV |
|
|
|
|
|
| 1000 mV, 10 kHz | ± 12.5 mV | ± 6.5 mV |
|
|
|
|
|
| 1000 mV, 20 kHz | N/A | ± 11.5 mV |
|
|
|
|
|
| 1000 mV, 30 kHz | ± 20.0 mV | N/A |
|
|
|
|
|
| 1000 mV, 100 kHz | N/A | ± 47 mV |
|
|
|
|
|
|
|
|
|
Agilent U1251A/U1252A User’s and Service Guide | 123 |