Agilent Technologies Frequency Sensitivity Specifications, Model Number U1251A U1252A, 100 kHz

Models: U1251A

1 185
Download 185 pages 44.17 Kb
Page 178
Image 178

8 Specifications

Frequency Sensitivity Specifications

For voltage measurements

Table 7-12Frequency sensitivity and trigger level specifications for voltage measurements

 

Minimum sensitivity (RMS sine wave)

 

Trigger level for DC coupling

 

 

 

 

 

 

 

 

 

 

 

 

Input

 

 

 

Model Number

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

U1251A

U1252A

U1251A

U1252A

range [1]

 

20 Hz - 100

>100 kHz -

20 Hz - 200

>200 kHz -

< 100 kHz

 

>100 kHz -

< 100 kHz

 

> 100 kHz -

 

kHz

200 kHz

kHz

500 kHz

 

200 kHz

 

500 kHz

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

50.000 mV

10 mV

15 mV

10 mV

25 mV

10 mV

 

15 mV

10 mV

 

25 mV

 

 

 

 

 

 

 

 

 

 

 

500.00 mV

25 mV

35 mV

70 mV

150 mV

60 mV

 

70 mV

70 mV

 

150 mV

 

 

 

 

 

 

 

 

 

 

 

1000.0 mV

40 mV

50 mV

120 mV

300 mV

100 mV

 

150 mV

120 mV

 

300 mV

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.5 V/1.25

 

 

 

 

 

5.0000 V

0.25 V

0.5 V

0.3 V

1.2 V

V (< 100

 

0.6 V

0.6 V

 

1.5 V

 

 

 

 

 

Hz)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

50.000 V

2.5 V

5 V

3 V

5 V

5 V

 

6 V

6 V

 

15 V

 

 

 

 

 

 

 

 

 

 

 

500.00 V

25 V

-

30 V < 100

-

50 V

 

-

60 V

 

-

kHz

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1000.0 V

50 V

-

50 V < 100

-

300 V

 

-

120 V

 

-

kHz

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Notes for frequency sensitivity and trigger level specifications for voltage measurements;

 

 

 

1Maximum input for specified accuracy = 10 x range or 1000 V.

2The input signal is lower than the product of 20,000,000 V-Hz.

158