7 Performance Tests and Calibration
Step |
| Test Function | Cal Range | Input |
| Cal Item |
| |
|
|
|
|
| ||||
| U1251A |
| U1252A | |||||
|
|
|
|
|
| |||
|
|
|
|
|
|
|
| |
7 | Turn the rotary switch to the | Open | Input terminal | oPEn |
| oPEn |
| |
| TEMP/ | position |
| open (Remove any |
|
|
|
|
|
|
|
| test leads and |
|
|
|
|
|
|
|
| Short Plugs from |
|
|
|
|
|
|
|
| the input terminal) |
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
| 10 nF | 3 nF | 03.000 nF |
| 03.000 nF | |
|
|
|
| 10 nF | 10.000 nF |
| 10.000 nF | |
|
|
|
|
|
|
|
| |
|
|
| 100 nF | 10 nF | 010.00 nF |
| 010.00 nF | |
|
|
|
| 100 nF | 100.00 nF |
| 100.00 nF | |
|
|
|
|
|
|
|
| |
|
|
| 1000 nF | 100 nF | 0100.0 nF |
| 0100.0 nF | |
|
|
|
| 1000 nF | 1000.0 nF |
| 1000.0 nF | |
|
|
|
|
|
|
|
|
|
|
|
| 10 μF | 10 μF | 10.000 μF |
| 10.000 | μF |
|
|
|
|
|
|
|
|
|
|
|
| 100 μF | 100 μF | 100.00 μF |
| 100.00 | μF |
|
|
|
|
|
|
|
|
|
|
|
| 1000 μF | 1000 μF | 1000.0 μF |
| 1000.0 | μF |
|
|
|
|
|
|
|
| |
|
|
| 10 mF | 10 mF | 10.000 mF |
| 10.000 mF | |
|
|
|
|
|
|
|
|
|
8 | Press | button to go to | N/A | 0 ºC | 0000.0 ºC |
| 0000.0 | ºC |
| TEMP mode |
|
|
|
|
|
| |
|
|
|
|
|
| |||
| Note: Ensure the multimeter is turned on and stabilized for at least 60 minutes, with the | |||||||
| connected between the multimeter and the calibrator output terminal |
|
|
|
| |||
|
|
|
|
|
|
|
| |
9 | Turn the rotary swich to the | OPEN | Input terminal | oPEn |
| oPEn |
| |
|
| mA position |
| open (Remove any |
|
|
|
|
|
|
|
| test leads and |
|
|
|
|
|
|
|
| Short Plugs from |
|
|
|
|
|
|
|
| the input terminal) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 500 μA | 300 μA | 300.00 μA |
| 300.00 | μA |
|
|
|
|
|
|
|
|
|
|
|
| 5000 μA | 3000 μA | 3000.0 μA |
| 3000.0 | μA |
|
|
|
|
|
|
|
|
|
140 | Agilent U1251A/U1252A User’s and Service Guide |