Performance Tests and Calibration

7

 

 

 

 

 

 

 

 

 

 

Step

 

 

 

Test Function

Range

5520A Output

Error from nominal 1 year

 

 

 

 

 

 

 

 

 

U1251A

U1252A

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

100.00 μF

100.00 μF

± 1.05 μF

± 1.05 μF

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1000.0 μF

1000.0 μF

± 10.5 μF

± 10.5 μF

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

10.00 mF

10.00 mF

± 0.105 mF

± 0.105 mF

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

100.00 mF

10.00 mF

± 0.4 mF

± 0.4 mF

 

 

 

 

 

 

 

 

 

 

 

14

Press

 

button to go to

–200 ºC until

0 ºC

± 3 ºC

± 3 ºC

 

 

TEMP mode [8][13]

1372 ºC

 

 

 

 

 

 

 

 

 

 

100 ºC

± 3.3 ºC

± 3.3 ºC

 

 

 

 

 

 

 

 

 

 

15

Turn the rotary swich to the

500 μA

500 μA

± 0.55 μA [9]

± 0.3 μA [9]

 

 

 

 

uA position

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5000 μA

5000 μA

± 5.5 μA [9]

± 3 μA [9]

 

16

Press

 

button to go to

500 μA

500 μA, 1 kHz

± 4.2 μA

± 3.7 μA

 

 

 

 

 

uA mode [1]

 

500 μA, 20 kHz

± 15.8 μA

± 3.95 μA

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5000 μA

5000 μA, 1 kHz

± 42 μA

± 37.0 μA

 

 

 

 

 

 

 

5000 μA, 20 kHz

± 0.156 mA

± 39.5 μA

 

 

 

 

 

 

 

 

 

 

17

Turn the rotary switch to the

50 mA

50 mA

± 0.105 mA [9]

± 80 μA [9]

 

 

 

 

mA.A position

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

440 mA

400 mA

± 0.93 mA [9]

± 0.71 mA [9]

 

18

Press

 

button to go to

50 mA

50 mA, 1 kHz

± 0.42 mA

± 0.37 mA

 

 

 

 

 

mA mode [1]

 

50 mA, 20 kHz

± 1.56 mA

± 0.395 mA

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

440 mA

400 mA, 45 Hz

± 3.4 mA

± 3.0 mA

 

 

 

 

 

 

 

400 mA, 1 kHz

± 3.4 mA

± 3.0 mA

 

 

 

 

 

 

 

 

 

 

 

Caution: Connect calibrator to hanheld multimeter's A and COM terminal before applying 5A and 10A

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5 A

5 A

± 16 mA

± 16 mA

 

 

 

 

 

 

10 A [10]

10 A

± 40 mA

± 35 mA

 

 

 

 

 

 

 

 

 

 

 

Agilent U1251A/U1252A User’s and Service Guide