2 Making Measurements
Audible continuity
In the range of
Table 2-3 Audible continuity measurement range
Measurement range | Beeper sound threshold |
|
|
500.00 Ω | < 10 Ω |
|
|
5.0000 kΩ | < 100 Ω |
|
|
50.000 kΩ | < 1 kΩ |
|
|
500.00 kΩ | < 10 kΩ |
|
|
5.0000 MΩ | < 100 kΩ |
|
|
50.000 MΩ | < 1 MΩ |
|
|
500.00 MΩ | < 10 MΩ |
|
|
Conductance
Set up the multimeter to measure conductance as shown in
Figure 2- 9. Probe the test points and read the display.
The conductance measurement enables the measurement of very high resistance of up to 100 GΩ.
As the high- resistance readings are susceptible to noise, you can capture the average readings by using the Dynamic Recording mode. Refer to the section “Dynamic Recording” on page 50 for more information.
36 | Agilent U1251A/U1252A User’s and Service Guide |