Agilent Technologies U1251A manual 7U1252A true RMS ac+dc current specifications

Models: U1251A

1 185
Download 185 pages 44.17 Kb
Page 175
Image 175

 

 

 

 

 

 

Specifications

8

AC+DC Specifications for U1252A

 

 

 

 

 

Table 7-6U1252A true RMS ac+dc voltage specifications

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Frequency

 

 

 

 

 

 

30 Hz

45 Hz

1 kHz

10 kHz

20kHz

 

Function

Range

Resolution

45 Hz

1 kHz

10 kHz

20 kHz

100kHz [1]

 

 

50.000 mV

0.001 mV

1.5 + 80

0.4 + 60

0.7 + 60

0.8 + 60

3.5 + 220

 

 

500.00 mV

0.01 mV

1.5 + 65

0.4 + 30

0.4 + 30

0.8 + 45

3.5 + 125

 

 

1000.0 mV

0.1 mV

1.5 + 65

0.4 + 30

0.4 + 30

0.8 + 45

3.5 + 125

 

Voltage [2]

5.0000 V

0.0001 V

1.5 + 65

0.4 + 30

0.4 + 30

0.8 + 45

3.5 + 125

 

 

50.000 V

0.001 V

1.5 + 65

0.4 + 30

0.4 + 30

0.8 + 45

3.5 + 125

 

 

500.00 V

0.01 V

1.5 + 65

0.4 + 30

0.4 + 30

1.5 + 45

3.5 + 125 [3]

 

 

1000.0 V

0.1 V

1.5 + 65

0.4 + 45

0.4 + 45

1.5 + 45 [3]

-

 

Notes for U1252A AC +DC voltage specifications:

1The additional error to be added as frequency > 20 kHz and signal input <10 %of range: 3 counts of LSD per kHz.

2Input impedance: > 1 GW for 50 mV to 1000 mV. Input impedance is 1.1 MΩ (nominal) with <100 pF for 5 V to 1000 V ranges.

3The input signal is lower than the product of 20,000,000 V-Hz (product of voltage and frequency).

Table 7-7U1252A true RMS ac+dc current specifications

 

 

 

 

Frequency

 

Function

Range

Resolution

30 Hz – 45 Hz

45 Hz – 1 kHz

1 kHz – 20 kHz

 

500.00 µA [1]

0.01 µA

1.1 + 25

0.8 + 25

0.8 + 25

 

5000.0 µA

0.1 µA

1.1 + 25

0.8 + 25

0.8 + 25

 

 

 

 

 

 

Current

50.000 mA

0.001 mA

1.2 + 25

0.9 + 25

0.9 + 25

 

 

 

 

 

440.00 mA

0.01 mA

1.2 + 25

0.9 + 25

0.9 + 25

 

 

 

 

 

 

 

 

5.0000 A

0.0001 A

1.8 + 30 [3]

0.9 + 30

3.3 + 70

 

10.000 A [2]

0.001 A

1.8 + 30 [3]

0.9 + 25

< 3 A/5 kHz

Notes for U1252A AC +DC current specifications:

1Input current > 35 µArms.

2Current can be measured from 2.5 A up to 10 A continuously. Add 0.5% to the specified accuracy if the signal measured is in the range of 10 A to 20 A and for a period of up to 30 seconds. After measuring current > 10 A, leave the meter to cool down for a period that is twice the measuring time used before application of low current measurement.

3Input current < 3 Arms.

Agilent U1251A/U1252A User’s and Service Guide