7 Performance Tests and Calibration

Step

 

Test Function

Cal Range

Input

 

Cal Item

 

 

 

 

 

U1251A

 

U1252A

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3

Press

button to go to the

5 V

0.3 V, 1 kHz

N/A

 

0.3000 V

 

 

V mode

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3 V, 1 kHz

N/A

 

3.0000 V

 

 

 

 

3 V, 50 kHz

N/A

 

3.0000 V

 

 

 

 

 

 

 

 

 

 

 

50 V

3 V, 1 kHz

N/A

 

03.000 V

 

 

 

 

30 V, 1 kHz

N/A

 

30.000 V

 

 

 

 

30 V, 50 kHz

N/A

 

30.000 V

 

 

 

 

 

 

 

 

 

 

 

500 V

30 V, 1 kHz

N/A

 

030.00 V

 

 

 

 

300 V, 1 kHz

N/A

 

300.00 V

 

 

 

 

300 V, 50 kHz

N/A

 

300.00 V

 

 

 

 

 

 

 

 

 

 

 

1000 V

30 V, 1 kHz

N/A

 

0030.0 V

 

 

 

 

300 V, 1 kHz

N/A

 

0300.0 V

 

 

 

 

300 V, 50 kHz

N/A

 

0300.0 V

 

 

 

 

 

 

 

4

Turn the rotary switch to the

Short

Dual banana plug

SHort

 

SHort

 

 

mV position

 

with copper wire

 

 

 

 

 

 

 

short between 2

 

 

 

 

 

 

 

terminals

 

 

 

 

 

 

 

 

 

 

 

 

 

 

50 mV

30 mV

30.000 mV

 

30.000 mV

 

 

 

 

 

 

 

 

 

 

 

500 mV

300 mV

300.00 mV

 

300.00 mV

 

 

 

 

 

 

 

 

 

 

 

1000 mV

1000 mV

1000.0 mV

 

1000.0 mV

 

 

 

 

 

 

 

 

138

Agilent U1251A/U1252A User’s and Service Guide

Page 158
Image 158
Agilent Technologies manual Agilent U1251A/U1252A User’s and Service Guide