Performance Tests and Calibration

7

Table 6-2Performance verification test steps

Step

 

Test Function

Range

5520A Output

Error from nominal 1 year

 

 

 

 

U1251A

U1252A

 

 

 

 

 

 

 

 

 

 

 

1

Turn the rotary switch to the

5 V

5 V, 1 kHz

± 32.5 mV

± 22.5 mV

 

 

V position [1]

 

 

 

 

 

 

 

 

5 V, 10 kHz

± 52.5 mV

± 22.5 mV

 

 

 

 

5 V, 20 kHz

N/A

± 41.5 mV

 

 

 

 

5 V, 30 kHz

± 84 mV

N/A

 

 

 

 

4.5 V, 100 kHz

N/A

± 169.5 mV

 

 

 

 

 

 

 

 

 

 

50 V

50 V,1 kHz

± 325 mV

± 225 mV

 

 

 

 

50 V,10 kHz

± 525 mV

± 225 mV

 

 

 

 

50 V, 20 kHz

N/A

± 415 mV

 

 

 

 

50 V, 30 kHz

± 840 mV

N/A

 

 

 

 

45 V, 100 kHz

N/A

± 1.695 V

 

 

 

 

 

 

 

 

 

 

500 V

500 V, 1 kHz

± 3.25 V

± 2.25 V

 

 

 

 

500 V, 10 kHz

± 5.25 V

± 2.25V

 

 

 

 

 

 

 

 

 

 

1000 V

1000 V, 1 kHz

± 10 V

± 8.0 V

 

 

 

 

 

 

 

2

Press

button to go to

9.9999 kHz

0.48 V, 1 kHz

± 500 mHz

± 500 mHz

 

frequency mode

 

 

 

 

 

 

 

 

 

 

 

3

Press

button to go to

0.01% – 99.99%

5.0 Vpp @

± 0.315%

± 0.315%

 

Duty Cycle mode

 

50%, Square

 

 

 

 

 

 

Wave, 50 Hz

 

 

 

 

 

 

 

 

 

Agilent U1251A/U1252A User’s and Service Guide

121

Page 141
Image 141
Agilent Technologies U1251A 2Performance verification test steps, Step Test Function Range, Error from nominal 1 year