5 Maintenance
4To test Fuse 2, place and touch the tip of the test probe on the right half of input terminal . Ensure that the probe tip touches the metal inside the input terminal.
5Observe the reading on the instrument's display. Refer to Table 5- 3 for the possible readings that could appear).
6Replace the fuse when OL is displayed.
Table 5-3 Measurement readings for fuse checking
Current input terminal | Fuse | Fuse rating | Fuse OK (approximately) | Replace fuse | |
|
| ||||
Displayed readings | |||||
|
|
| |||
|
|
|
|
| |
| 1 | 440 mA/1000 V | 6.2 MΩ | OL | |
|
|
|
|
| |
| 2 | 11 A/1000 V | 0.06 Ω | OL | |
|
|
|
|
|
Replacing the fuse
N O T E
This manual provides only the fuse replacement procedures, but not the fuse replacement markings.
Follow the procedures below to replace the fuse.
1Turn the meter off and disconnect the test leads from the external equipment. Make sure to remove the adaptor.
2Wear clean/dry gloves and do not touch any other component except the fuse and plastic parts. The current calibration is considered shunt only, so do not recalibrate the meter after replacing the fuse.
3Remove the battery cover compartment to replace the fuse.
4Loosen the three screws on the bottom case and remove the cover.
5Loosen the two screws on the top corners to lift the circuit board.