Specifications 8
Electrical Specifications
DC Specifications
Table
|
|
| Test Current/ |
| Accuracy | |
Function | Range | Resolution | Burden Voltage | U1251A |
| U1252A |
| 50.000 mV | 0.001 mV | - | 0.05 + 50 [2] | 0.05 + 50 [2] | |
| 500.00 mV | 0.01 mV | - |
|
|
|
| 1000.0 mV | 0.1 mV | - |
|
|
|
Voltage [1] | 5.0000 V | 0.0001 V | - | 0.03 + 5 | 0.025 + 5 | |
| 50.000 V | 0.001 V | - |
|
| |
|
|
|
| |||
| 500.00 V | 0.01 V | - |
| 0.03 + 5 | |
| 1000.0 V | 0.1 V | - |
|
|
|
Notes for DC voltage specifications:
1Input impedance: >1GΩ for 50 mV to 1000 mV ranges. For U1251A, input impedance is 10 MΩ (nominal) for 5 V to 1000 V
ranges. For U1252A, input impedance is 10 MΩ (nominal) in parallel with 1.1 MΩ at dual display.
2The accuracy could be 0.05 % +10 for U1251A and 0.05 % + 5 for U1252A. Always use the Null function to zero out the thermal effect before measuring the signal.
|
|
| Test Current/ |
| Accuracy |
Function | Range | Resolution | Burden Voltage | U1251A | U1252A |
| 500.00 Ω [3] | 0.01 Ω | 1.04 mA | 0.08 + 10 | 0.05 + 10 |
| 5.0000 kΩ [3] | 0.0001 kΩ | 416 µA |
|
|
| 50.000 kΩ | 0.001 kΩ | 41.2 µA | 0.08 + 5 | 0.05 + 5 |
|
|
|
|
| |
| 500.00 kΩ | 0.01 kΩ | 4.12 µA |
| |
|
|
| |||
|
|
|
|
|
|
Resistance | 5.0000 MΩ | 0.0001 MΩ | 375 nA | 0.2 + 5 | 0.15 + 5 |
| 50.000 MΩ [4] | 0.001 MΩ | 187 nA | 1 + 10 | 1 + 5 |
| 500.00 MΩ [4] | 0.01 MΩ | 187 nA | - | 3 + 10 < 200MΩ/ |
| 8 + 10 > 200MΩ | ||||
|
|
|
|
| |
| 500.00 nS [5] | 0.01 nS | 187 nA | 1 + 20 | 1 + 10 |
Notes for resistance specifications:
3The accuracy of 500 Ω and 5 kΩ is specified after Null function, which is used to subtract the test lead resistance and the thermal effect.
4For the range of 50/500 MΩ, the R.H. is specified for < 60 %.
5The accuracy is specified for < 50 nS and after Null function with open test lead.
Agilent U1251A/U1252A User’s and Service Guide | 151 |