Performance Tests and Calibration | 7 |
Calibration interval
A 1- year interval is adequate for most applications. Accuracy specifications are warranted only if adjustment is made at regular calibration intervals. Accuracy specifications are not warranted beyond the 1- year calibration interval. Agilent does not recommend extending calibration intervals beyond 2 years for any application.
Adjustment is recommended
Specifications are only guaranteed within the period stated from the last adjustment. Agilent recommends that readjustment should be performed during the calibration process for best performance. This will ensure that the U1251A/U1252A will remain within the specifications for the next calibration interval. This criterion for the re- adjustment provides the best
Performance data are measured during the Performance Verification Tests but this does not guarantee that the instrument will remain within these limits unless the adjustments are performed.
Refer to the section “To read the calibration count” on page 142 and verify that all the adjustments have been performed.
Agilent U1251A/U1252A User’s and Service Guide | 113 |