8. Before calling the service technician (Warning)

module id message id

Message

Detail

 

 

 

 

19

Specified End Date is earlier than Start Date.

Input a date later than the Start Date as the End Date.

 

 

 

 

 

 

Length of acquired worklist item is too long.

 

20

The following tag number item(s) are cut off.

 

[XXXX] (DICOM TAG Number+DICOM TAG

 

 

 

The transfer data of the worklist server does not

10045

 

Name)

 

meet the DICOM standards. Check the worklist

 

 

 

 

 

Acquired worklist item format is invalid.

 

 

server settings.

 

 

The following tag number item(s) are not acquired.

21

 

[XXXX] (DICOM TAG Number+DICOM TAG

 

 

 

 

 

 

Name)

 

 

 

 

1

Invalid Values for Birth Date

Input the correct birth date.

 

 

 

 

5

Patient Info. set as any of essential data items are

Input the patient information that was set in the

not entered.

essential data (see page 83).

 

 

 

 

 

 

 

 

The length of either of 5 Patient Name fields is too

 

6

long.

Shorten the length to 60 characters or less.

 

 

Shorten the length to 60 characters or less.

 

 

 

 

 

12

Invalid character was entered to the field.

Check the details of the displayed item, and enter it

again.

 

 

 

 

 

 

 

14

Line feed code cannot be entered to Study

Delete the line feed code.

Description field.

 

 

 

22Character '\' or control character cannot be used for Accession No.

 

 

 

Delete the backslash (\) and any control characters.

 

23

Character '\' or control character cannot be used for

 

 

 

Patient ID.

 

 

 

 

 

 

 

 

 

24

Character '\,' line feed code, or control character

Delete the backslash (\), line feed code and any

 

cannot be used for Study Description.

control characters.

 

 

 

 

 

 

 

29

Character '\', '=' or control code cannot be used for

Delete the backslash (\) and any control characters.

 

 

Patient Name.

 

 

30

Multi-byte character cannot be used for Patient

Delete the multi byte characters.

 

Name.

 

 

 

 

 

 

 

10048

39

Out-of-Range Birth Date (A date earlier than 1869

 

 

or later than tomorrow is entered.)

Input the correct birth date.

 

 

 

40

Entered date is invalid for Birth Date.

 

42A character/characters for Patient ID is/are not supported by Specific Character Set.

43A character/characters for Accession No. is/are not supported by Specific Character Set.

44A character/characters for Patient Name is/are not supported by Specific Character Set.

45

A character/characters for Study Description is/are

Input using the specific character set.

 

not supported by Specific Character Set.

 

46A character/characters for the field is/are not supported by Specific Character Set.

47A character/characters for the field is/are not supported by Specific Character Set.

48A character/characters for the field is/are not supported by Specific Character Set.

49

Entered characters are not acceptable because they

Delete the single-byte Japanese kana characters.

 

do not match Specific Character Set.

 

50

Length of the acquired item is too long.

Check the information input in the magnetic card or

The following item(s) is/are cut off.

barcode.

 

-122-