Analog Programming (APG) and Isolated Analog Programming (ISOL)

Query for Analog Voltage Input Level

Quick Tip Remote operation uses these SCPI commands. See page 5–1.

To query for analog voltage input level from non-isolated input:

1.Turn the 9-position mode control to the VAP position or press the rotary Adjust/Enter control if the mode control is already at the VAP position.

UAPr is displayed on the output voltage display.

2.If the mode is currently nonE then select any other APG mode. If this step is done, be sure to restore the APG setting to nonE when the operation is complete. Otherwise, do not rotate the rotary Adjust/Enter control to keep the current APG mode.

3.Press the rotary Adjust/Enter control to select the APG mode.

The output voltage display should read LE U and the output current display will display APG level.

4.Press the rotary Adjust/Enter control to return to normal mode.

5.If the APG mode was modified in step 2, execute the procedure to deactivate the APG voltage mode.

The SCPI command for these instructions are:

[:]SYSTem[<channel>]:REMote:SOURce[:VOLTage] <?LOCal AVOLtageIAVoltageARESistiveIAResistive>

[:]SYSTem[<channel>]:COMMunicate:APRogram:LEVel[:VOLTage]

[:ISOlated]

[:]MEASure[<channel>][:SCALar]:APRogram[:VOLTage][:DC]?

4-16

M370078-01

Page 108
Image 108
Xantrex Technology XG 850 manual Query for Analog Voltage Input Level, Scpi command for these instructions are