Table A-4SCPI Commands for Calibration

 

 

 

Parameter and

 

Function

SCPI Command

Description

Range

Query

 

 

 

 

 

Restore

[:]CALibration[<channel>]:RESTore

Restores the calibration to

N/A

N/A

Factory

 

the factory defaults.

 

 

Calibration

 

 

 

 

 

 

 

 

 

Calibrate

[:]CALibration[<channel>][:VOLTage]

Calibrate the over voltage

<OVP Voltage Set

No

OVP

:PROTection[:OVER]

protection.

in Volts>

 

protection.

 

 

 

 

 

 

 

 

 

Enter Output

[:]CALibration[<channel>]:OUTPut

Sets the voltage output

<Voltage set in

No

Voltage Data

:VOLTage

calibration level.

volts>

 

 

 

 

 

 

Enter Output

[:]CALibration[<channel>]:OUTPut

Sets the current output

<Current set in

No

Current Data

:CURRent

calibration level

amps>

 

 

 

 

 

 

Calibrate the

[:]CALibration[<channel>]:MONitor

Calibrate the voltage

? <NR1>

Yes

Voltage

[:VOLTage]

monitor signal system.

<nothing for

 

controlled

 

 

interactive mode>

 

Non Isolated

 

 

 

 

Voltage APG

 

 

 

 

feature

 

 

 

 

 

 

 

 

 

Calibrate the

[:]CALibration[<channel>]:MONitor

Calibrate the isolated

? <NR1>

Yes

Voltage

[:VOLTage]:ISOLated

voltage monitor signal

<nothing for

 

controlled

 

system.

interactive mode>

 

Isolated

 

 

 

 

Voltage APG

 

 

 

 

feature

 

 

 

 

 

 

 

 

 

Calibrate the

[:]CALibration[<channel>]:MONitor

Calibrate the current

? <NR1>

Yes

Voltage

:CURRent

monitor signal system.

<nothing for

 

controlled

 

 

interactive mode>

 

Non Isolated

 

 

 

 

Current APG

 

 

 

 

feature

 

 

 

 

 

 

 

 

 

SCPI Command Reference

Page 244
Image 244
Xantrex Technology XG 850 manual Table A-4SCPI Commands for Calibration