Configure Other Protection Mechanisms |
| |
Configure Other Protection Mechanisms |
| |
Foldback Protection |
| |
Foldback protection causes the output of the power supply to shut down if |
| |
the selected regulation mode is entered and the configured delay time |
| |
expires. A delay time may be specified as well. The only way to clear |
| |
foldback is by pressing the rotary Adjust/Enter control for 3 seconds and |
| |
executing the Clear command. |
| |
Command: |
| |
[:]OUTPut[<channel>]:PROTection:CLEar |
| |
Setting the Foldback Mode |
| |
The following command will set the foldback mode. The mode selected |
| |
by the character mnemonic sent will indicate which mode will trigger the |
| |
foldback alarm to begin counting. For example, if the command is sent |
| |
with the CV mnemonic and the output is enabled with no load attached |
| |
(open circuit), the unit will operate in constant voltage mode and the |
| |
foldback counter will begin counting. If the CC mode was set and the |
| |
output was enabled with no load connected, the unit will operate in CV |
| |
mode and the foldback counter will not trigger. |
| |
To set the foldback mode, send the command: |
| |
| ||
Command: | 5 | |
[:]OUTPut[<channel>]:PROTection:FOLDback {?CCCVNONE} | ||
| ||
Where: |
|
NONE indicates foldback protection is disabled.
CCindicates the supply will shut down due to constant current condition. CV indicates the supply will shut down due to constant voltage.
The default value is NONE.
To set the foldback delay, send the command:
Command:
[:]OUTPut[<channel>]:PROTection:FOLDback:DELay <delay_time>
Where:
<delay_time> is a value in the range of
The default value is 0.5 seconds.