Calibration and Troubleshooting
Over Voltage Protection Calibration
Important: The Voltage Calibration must be done before performing this procedure. If this is not done the OVP calibration will be inaccurate.
To calibrate the over voltage protection:
1.Turn on the power supply.
2.Select the VAP mode to be Off.
3.Set the voltage set point to be 90% of the model voltage and the current to be 10% of the model current.
4.Turn on the main power output.
5.Execute the OVP calibration SCPI command:
:CAL:PROT <output value set in step 3>
6.The power supply will enter into
Calibration of the power supply is complete.
Important: The output (main) voltage must be precalibrated.
The SCPI command (s) for these instructions are:
[:]CALibration[:VOLTage]:PROTection[:OVER] {<expected output value>}