Analog Programming Mode
Query for Analog Voltage Input Level
Quick Tip Remote operation uses these SCPI commands. See page
To query for analog voltage input level from
1.Turn the
UAPr is displayed on the output voltage display.
2.If the mode is currently nonE then select any other APG mode. If this step is done, be sure to restore the APG setting to nonE when the operation is complete. Otherwise, do not rotate the rotary Adjust/Enter control to keep the current APG mode.
3.Press the rotary Adjust/Enter control to select the APG mode.
The output voltage display should read LE U and the output current display will display APG level.
4.Press the rotary Adjust/Enter control to return to normal mode.
5.If the APG mode was modified in step 2, execute the procedure to deactivate the APG voltage mode.
The SCPI commands for these instructions are: | 4 |
[:]SYSTem[<channel>]:REMote:SOURce[:VOLTage] <?LOCal AVOLtageIAVoltageARESistiveIAResistive>
[:]SYSTem[<channel>]:COMMunicate:APRogram:LEVel[:VOLTage]
[:ISOlated]
[:]MEASure[<channel>][:SCALar]:APRogram[:VOLTage][:DC]?