Remote Operation

Locking and Unlocking the Front Panel

Locking out the front panel will prevent any of the buttons from functioning. All the buttons and knobs on the front panel will display the LOCL Loc message to be display on the Current and Voltage displays if pressed or rotated. This mode prevents any changes to the unit from the front panel. See the following procedure to lock and unlock the front panel.

The SCPI Commands for these instructions are:

[:]SYSTem[<channel>]:REMote:STATe {?LOCalREMote}

Where:

LOCal is used to set the front panel mode to local.

REMote is used to set the front panel into locked mode (remote). For example:

SYST:REM:STAT REM Front panel is locked (remote digital programming mode only).

SYST:REM:STAT LOC Front panel is unlocked.

SYST:REM:STAT? Query for status

Auto Sequence Programming

Auto Sequence programming lets you execute a program stored in the power supply’s memory. The program that can be stored in memory can be composed of any of the SCPI commands described in this Manual.

For example, Auto Sequence programming allows you to set the timer to turn the power output of the supply Off or On after a time interval or to program voltage sequences.

When a program is activated each command stored in the buffer is executed as if the operator had sent the command by typing it in. A variable delay time called dwell is spent between commands. The dwell time can range from 0 seconds to 180 seconds. See “Setting Dwell Time” on page 5–61.

The Auto Sequence program will be cleared after AC power is turned off; however, a procedure exists for storing it in a text file which can be reloaded at a later date.

5-60

M370078-01

Page 186
Image 186
Xantrex Technology XG 850 manual Locking and Unlocking the Front Panel, Auto Sequence Programming