Local Lockout
Local Lockout
Local lockout is a feature that allows the front panel to be locked so that accidental button presses are ignored. This feature is often used to lockout the front panel when you are controlling the power supply from a remote location. When in local lockout mode, the front panel will display LOCL Loc whenever a button is pressed or a knob is turned.
Enabling Local Lockout
To enable local lockout:
1.Turn the
The display will show rE in the output voltage display and the | 3 |
| |
selected remote interface in the output current display. |
|
2.Turn the rotary Adjust/Enter control until the display shows LOCL Loc.
3.Press the rotary Adjust/Enter control.
4.Turn the rotary Adjust/Enter control until the display shows LOCL On.
5.Press the rotary Adjust/Enter control. Local lockout has been enabled.
The SCPI command (s) for these instructions are:
[:]SYSTem:REMote:STATe
Disabling Local Lockout
To disable local lockout:
1.Turn the
The display will show rE in the output voltage display and the selected remote interface in the output current display.
2.Turn the rotary Adjust/Enter control until the display shows LOCL Off.
3.Press the rotary Adjust/Enter control. Local lockout has been disabled.