Isolated Current Monitoring Calibration

Important: For maximum accuracy at a specific APG level. Follow the calibration procedure again but use the intended operating voltage level in steps 3 and 8.

To calibrate the isolated current monitoring:

1.Short the main output with a shunt. Connect a multimeter to measure the voltage across the shunt.

2.Connect the voltmeter to the Isolated monitoring output lines (J3.10 and J3.6).

3.Set the current APG level to 4 V. The SCPI command is:

SYST:COMM:APR:LEV:CURR 4

4.Set the input source to local for voltage analog programming. the SCPI command is:

SYST:REM:SOUR:CURR LOC

5. Set the main output voltage to 100% by sending the command:

SOUR:VOLT <ModelVolt>

or use the rotary Adjust/Enter control.

6.Turn the main output to On by pressing the OUTPUT ENABLE Main button on the front panel or send the command:

 

:OUTP ON

 

 

The OUTPUT ENABLE Main button will illuminate.

 

7.

Enter the interactive calibration mode by sending the SCPI command:

 

 

CAL:MON:CURR:ISOL

 

 

 

8.

Press the terminal keys – or + to reduce or increase the monitoring

6

 

value. Adjust the monitoring value displayed on the voltmeter that

 

 

 

was set up in Step 1 to read as close as possible to 4.000 V.

 

9.

Press any key other than the + or – keys to exit from interactive

 

 

calibration mode.

 