Isolated Analog Programming Mode (ISOL)
M370078-01 4-31
4
Resistive-Controlled Current ISOL Setup

Activating ISOL Resistive-Controlled Current Setup

1. Turn the 9-position mode control to the CAP position or press the
rotary Adjust/Enter control if the control knob is already at the CAP
position.
CAPr is displayed on the output voltage display.
2. Turn rotary Adjust/Enter control to select R IS (voltage source
from isolated connector).
3. Press the rotary Adjust/Enter control to commit the selected mode.
LE C is displayed on the output voltage display.
4. Select the input source level using the rotary Adjust/Enter control.
5. Once the desired level has been selected, press the rotary Adjust/
Enter control to commit the change.
The display will return to the normal display.

Deactivating ISOL Current Mode

To deactivate the ISOL current mode:
1. Turn the 9-position mode control to the CAP position or press the
rotary Adjust/Enter control if the control knob is already at the CAP
position.
2. CAPr is displayed on the output voltage display.
3. Turn the rotary Adjust/Enter control to select nonE.
4. Press the rotary Adjust/Enter control to commit the change.
The display will return to the normal display.