Reference: Remote Commands ~ Command Description

:SENSe:MEASure:APSTime:DETEcted

This query returns one or more values to indicate the receiver events detected during the most recently performed APS measurement.

Syntax: :SENSe<slot>:MEASure:APSTime:DETEcted?

Parameters

Description

NONE

No conditions were detected

LOS

Loss of Signal

OOF

Severely Errored Frame / Out of Frame

B1BIP

B1 errors

B2BIP

B2 errors

B3BIP

B3 errors

BIT

Bit (payload) errors

AIS_L

Line AIS / MS-AIS

AIS_P

Path AIS

RDI_L

Line RDI / MS-RDI

RDI_P

Path RDI

:SENSe:MEASure:APSTime:MAXTime

This command sets the maximum time for APS switch time measurement. The query form returns the current setting

Syntax: :SENSe<slot>:MEASure:APSTime:MAXTime? :SENSe<slot>:MEASure:APSTime:MAXTime <NR1>

:SENSe:MEASure:APSTime:RUN

This command starts (1) an APS switch time measurement, or stops (0) any measurement in progress. Syntax: :SENSe<slot>:MEASure:APSTime:RUN <Boolean>

:SENSe:MEASure:APSTime:STATus

This query returns the status of the most recently performed APS switch time measurement. Syntax: :SENSe<slot>:MEASure:APSTime:STATus?

Parameters

Description

INVALID

No valid result is available

BUSY

An APS time measurement is in process

PASS

The current APS time test passed

FAIL

The current APS time test failed

3-34

OTS9100 User Manual

Page 165
Image 165
Tektronix 071-0855-05 user manual SENSeMEASureAPSTimeDETEcted, SENSeMEASureAPSTimeMAXTime, SENSeMEASureAPSTimeRUN