Reference: Remote Commands ~ Command Description

:SENSe:MEASure:SECTion:WINDow:ALL

This query returns the values of all of the parameters of the :SENSE:MEASURE:SECTION:WINDOW? query, as measured during the most recent N-second period. The parameters are returned in the order in which they are listed under :SENSE:MEASURE:SECTION:CUMULATIVE.

Syntax :SENSe<slot>:MEASure:SECTion:WINDow:ALL?

:SENSe:MEASure:WINDow:CLEAR

This command clears the data in the sliding measurement window.

Syntax :SENSe<slot>:MEASure:WINDow:CLEAR

:SENSe:MEASure:WINDow:LEVEL

This query returns the number of seconds of measurement data included in the current windowed- measurement calculations. Normally this value will be the same as the window size; clearing the window data or changing the window size causes the sliding window to refill as new measurements are made.

Syntax :SENSe<slot>:MEASure:WINDow:LEVEL?

:SENSe:MEASure:WINDow:SIZE

This command sets the duration in seconds of the interval over which windowed results measurements are calculated. The query form returns the current setting.

Syntax :SENSe<slot>:MEASure:WINDow:SIZE? :SENSe<slot>:MEASure:WINDow:SIZE <NR1>

:SENSe:OPTical:INPut:OVERload

This query returns a Boolean value indicating whether or not the optical power has exceeded its maximum threshold. A return value of 1 indicates that optical power is overloaded, while a return value of 0 indicates that optical power is at an acceptable level. This command and query may be directed either to the OTS Optics card or to the associated Receiver card.

Syntax :SENSe<slot>:OPTical:INPut:OVERload?

3-40