Reference: Remote Commands ~ Command Description
3-40 OTS9100 User Manual
:SENSe:MEASure:SECTion:WINDow:ALL
This query returns the values of all of the parameters of the :SENSE :MEASURE:SECTION:WINDOW?
query, as measured during the most recent N-second period. The param eters are returned in the order in
which they are listed under :SENSE:MEASURE:SECTION:CUMULATIVE.
Syntax :SENSe<slot>:MEASure:SECTion:WINDow:ALL?
:SENSe:MEASure:WINDow:CLEAR
This command clears the data in the sliding measurement window.
Syntax :SENSe<slot>:MEASure:WINDow:CLEAR
:SENSe:MEASure:WINDow:LEVEL
This query returns the number of seconds of measurement data included in the current windowed-
measurement calculations. Normally this value will be the same as the wind ow size; clearing the window
data or changing the window size causes the sliding window to refill as ne w measurements are made.
Syntax :SENSe<slot>:MEASure:WINDow:LEVEL?
:SENSe:MEASure:WINDow:SIZE
This command sets the duration in seconds of the interval over which win dowed results measurements
are calculated. The query form returns the current setting.
Syntax :SENSe<slot>:MEASure:WINDow:SIZE?
:SENSe<slot>:MEASure:WINDow:SIZE <NR1>
:SENSe:OPTical:INPut:OVERload
This query returns a Boolean value indicating whether or not the optical power has exceeded its
maximum threshold. A return value of 1 indicates that optical power is overloaded, while a return value of
0 indicates that optical power is at an acceptable level. T his command and query may be directed either
to the OTS Optics card or to the associated Receiver card.
Syntax :SENSe<slot>:OPTical:INPut:OVERload?