Reference: Remote Commands ~ Command Description

:SENSe:TEST:TIME:ELAPsed

This query returns the time since the start of the current test.

Syntax :SENSe<slot>:TEST:TIME:ELAPsed?

:SENSe:TRIGger:MODE

This command sets the condition that will produce a trigger output from the OTS-9000 receiver. The query form returns the current setting.

Syntax :SENSe<slot>:TRIGger:MODE? :SENSe<slot>:TRIGger:MODE <Enum>

Parameters

Description

OFF

The trigger output is disabled

ANOMaly

A trigger occurs when an anomaly is received

DEFEct

A trigger occurs when a defect is received

PAYload

A trigger occurs when a payload data error is received

AD

A trigger occurs when an anomaly or defect is received

AP

A trigger occurs when an anomaly or payload data error is received

DP

A trigger occurs when a defect or payload data error is received

ADP

A trigger occurs when an anomaly, defect, or payload data error is received

:SOURce:CLOCk:SOURce

This command selects the timing source for the signal generated by the OTS-9000 transmitter. The query form returns the current setting.

Syntax :SOURce<slot>:CLOCk:SOURce? :SOURce<slot>:CLOCk:SOURce <Enum>

Parameters

Description

INTErnal

Transmitter timing is based on the OTS-9000's internal reference

EXTErnal

Transmitter timing is derived from an external signal applied to the clock/trigger card

:SOURce:DATA:BACKground:STRUcture

This command sets the background structure of the signal generated by the OTS-9000 transmitter. The query form returns the current setting.

Syntax: :SOURce<slot>:DATA:BACKground:STRUcture? :SOURce<slot>:DATA:BACKground:STRUcture <Enum>

Parameters

Description

G10C

Bulk data at 10 Gb/s

G2P5C

Bulk data at 2.5 Gb/s

M622C

Bulk data at 622 Mb/s

M155C

Bulk data at 155 Mb/s

M51A

Bulk data at 51 Mb/s (SONET only)

OTS9100 User Manual

3-45

Page 176
Image 176
Tektronix 071-0855-05 user manual SENSeTESTTIMEELAPsed, SENSeTRIGgerMODE, SOURceCLOCkSOURce, SOURceDATABACKgroundSTRUcture