Reference: Remote Commands ~ Command Description

:SENSe:DATA:PATH:TRACe:TYPE

This query returns the most recently received J1 path trace type (length). See the command :SOURce:DATA:PATH:TRACe:TYPE for the format of the enum value returned.

Syntax: :SENSe<slot>:DATA:PATH:TRACe:TYPE?

:SENSe:DATA:PATH:TRACe:VALUe

This query returns the most recently received J1 path trace data. The data bytes are returned in IEEE- 488 block data format.

Syntax: :SENSe<slot>:DATA:PATH:TRACe:VALUe?

:SENSe:DATA:PATH:TRACe:VALUe:STRIng

This query returns the most recently received J1 path trace data. The data is returned as an ASCII string. Syntax: :SENSe<slot>:DATA:PATH:TRACe:VALUe:STRIng?

:SENSe:DATA:PAYLoad:PATTern

This command selects the test pattern for the received signal. The query form returns the current setting.

Syntax: :SENSe<slot>:DATA:PAYLoad:PATTern? :SENSe<slot>:DATA:PAYLoad:PATTern <Enum>

Parameters

Description

PRBS15A

ITU standard PRBS 2^15-1 pattern

PRBS15I

Inverted PRBS 2^15-1 pattern

PRBS23A

ITU standard PRBS 2^23-1 pattern

PRBS23I

Inverted PRBS 2^23-1 pattern

PRBS31A

ITU standard PRBS 2^31-1 pattern

PRBS31I

Inverted PRBS 2^31-1 pattern

ZEROs

Constant zero data

ONEs

Constant one data

USER

Constant 8-bit user-specified

LIVE

Live data

OTS9100 User Manual

3-29

Page 160
Image 160
Tektronix 071-0855-05 user manual SENSeDATAPATHTRACeTYPE, SENSeDATAPATHTRACeVALUeSTRIng, SENSeDATAPAYLoadPATTern