allowable

Apex 1 to 5.5 kW Generator

Table 4-6.Host Port Commands for RS-232 with AE Bus (Continued)

Command

 

 

Number of

Number of

 

 

Description

Host Data

Response

 

 

 

Bytes

Data Bytes

 

 

 

 

16

Specifies a percentage that the generator can

1 data byte

1

allowable

be out of setpoint before it starts the out-of-

8-bit value

 

deviation

setpoint timer. Accepts a value of 1 to 99%.

 

 

 

(Readback command is 185.)

 

 

 

 

 

 

19

Specifies the number of recipe steps. Send 1

1 data byte

 

number of

data byte that indicates the number of recipe

8-bit value

 

recipe

steps (0 through 5).

 

 

steps

 

 

 

 

 

 

 

 

22

Specifies setpoint for a recipe step. Send 3

3 data bytes

 

recipe step/

data bytes.

8-bit value

 

setpoint

byte 1 = recipe step number (can be 1

16-bit value

 

 

 

 

 

 

through 7)

 

 

 

bytes 2 and 3 = setpoint value (send least

 

 

 

 

significant byte first); the value must be

 

 

 

 

within the operating range of the Apex

 

 

 

 

unit (see “Output Electrical

 

 

 

 

Specifications” on page 3-6)

 

 

 

 

 

 

23

Sets the run time for the specified recipe step.

3 data bytes

 

recipe step/

Send 3 data bytes.

8-bit value

 

run time

byte 1 = recipe step number (1 through

16-bit value

 

 

 

 

 

 

7)

 

 

 

bytes 2 and 3 = run time in hundredths of

 

 

 

 

seconds or in joules (if the recipe is set

 

 

 

 

for joules mode; see CMD 28 to set

 

 

 

 

recipe type); send least significant byte

 

 

 

 

first

 

 

 

 

 

 

28

Sets the recipe for time or joules mode. Send

1 data byte

 

recipe type

one data byte indicating the recipe type:

8-bit value

 

 

1 = time mode

 

 

 

2 = joules mode

 

 

 

 

 

 

40

Sets the Host port timeout value. Accepts a

2 data bytes

1

host port

value of 2 to 500, representing 0.02 to 5.0 s.

16-bit value

 

timeout

(Readback command is 140.)

 

 

value

 

 

 

 

 

 

 

 

 

5708009-C

4-47