Conductivity/Resistivity Analyzer/Controller

8.3 Entering the Cal Factor for Each Cell

Introduction

Each type of cell has an associated cell constant; this number is part of the cell model number. The constant for each cell is entered during I/O setup. However, for greater precision, every Honeywell cell is individually tested at the factory, and a calibration factor unique to that cell is determined. The cal factor for a cell can be found on the plastic tag hanging from the cell lead wires.

ATTENTION

Any time you replace a cell, even with another cell of the same type, enter the cal factor for the replacement cell.

Procedure

To enter the cal factor for each cell, follow the procedure in Table 8-1.

Table 8-1 Procedure for Entering Cal Factor

Step

1

2

3

Screen

MENU

CONFIGURATION

CALIBRATION

MAINTENANCE

I/O SETUP

DIAGNOSTICS

CALIBRATION

CELL CAL FACTOR

CAL TRIM 1

CAL TRIM 2

REMOVE CAL TRIM C1

REMOVE CAL TRIM C2

DIAGNOSTICS

CELL CONSTANTS

 

C1

10.0

 

 

C2

10.0

 

 

 

 

 

DIAGNOSTICS

 

 

NEXT

 

 

 

 

 

 

 

 

 

EDIT CELL CAL FACTOR

 

 

C1

10.101

 

 

C2

09.955

 

DIAGNOSTICS

DPP <-- -->

Action

Go to CALIBRATION CELL CAL FACTOR.

The cell constant entered for each cell during I/O setup will be displayed.

To enter the unique cal factor for each cell, use the indicated function key to go into edit mode.

The current cell constant for each cell will be displayed. This information is read-only.

To change the values, use the function key labeled “NEXT”. To exit, use the MENU key.

For each cell, enter the cal factor found on the plastic tag hanging from the cell leads.

8-4

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

7/99

Page 102
Image 102
Honeywell 9782 Series manual Entering the Cal Factor for Each Cell, Procedure for Entering Cal Factor, Cell CAL Factor