
Conductivity/Resistivity Analyzer/Controller
8.3 Entering the Cal Factor for Each Cell
Introduction
Each type of cell has an associated cell constant; this number is part of the cell model number. The constant for each cell is entered during I/O setup. However, for greater precision, every Honeywell cell is individually tested at the factory, and a calibration factor unique to that cell is determined. The cal factor for a cell can be found on the plastic tag hanging from the cell lead wires.
ATTENTION
Any time you replace a cell, even with another cell of the same type, enter the cal factor for the replacement cell.
Procedure
To enter the cal factor for each cell, follow the procedure in Table
Table 8-1 Procedure for Entering Cal Factor
Step
1
2
3
Screen
MENU
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP
DIAGNOSTICS
CALIBRATION
CELL CAL FACTOR
CAL TRIM 1
CAL TRIM 2
REMOVE CAL TRIM C1
REMOVE CAL TRIM C2
DIAGNOSTICS
CELL CONSTANTS
| C1 | 10.0 |
|
| C2 | 10.0 |
|
|
|
| |
| DIAGNOSTICS |
| |
| NEXT |
| |
|
|
| |
|
|
|
|
| EDIT CELL CAL FACTOR |
| |
| C1 | 10.101 |
|
| C2 | 09.955 |
|
DIAGNOSTICS
DPP
Action
Go to CALIBRATION CELL CAL FACTOR.
The cell constant entered for each cell during I/O setup will be displayed.
To enter the unique cal factor for each cell, use the indicated function key to go into edit mode.
The current cell constant for each cell will be displayed. This information is
To change the values, use the function key labeled “NEXT”. To exit, use the MENU key.
For each cell, enter the cal factor found on the plastic tag hanging from the cell leads.
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual | 7/99 |