I/O Setup and System Configuration
Hold mode
During AutoClean operations alarms and outputs are always held. This means that no alarms will become active during the automatic operation. If one or more relays are being used for alarm annunciation, they will go to the
At the completion of the AutoClean operation the hold will be released and normal operation will resume after the configured delay.
Procedure
If AutoClean was selected during I/O setup, follow the procedure in Table
Table 5-8 Procedure for Configuring AutoClean
Step
1
2
Screen
MENU
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP
DIAGNOSTICS
PREV NEXT
CONFIGURATION
CAT/RETRANSMISSION
ALARMS
DISCRETE CONTROL
ADVANCED FEATURES
DIAGNOSTICS
PREV NEXT
ADVANCED FEATURES
CLOCK SETUP
COMPUTED VALUE
AUTO CLEAN
DIAGNOSTICS
PREV NEXT
ATTENTION
24 HOUR CLOCK MUST
BE SETUP FOR THIS FEATURE
DIAGNOSTICS
NEXT | |
|
|
Action
Go to CONFIGURATION ADVANCED FEATURES
AUTO CLEAN.
If you have already set the clock, go to the next screen.
If you have not set the clock, press MENU to exit this function, then set the clock as described in
7/99 | 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual |