I/O Setup and System Configuration

Hold mode

During AutoClean operations alarms and outputs are always held. This means that no alarms will become active during the automatic operation. If one or more relays are being used for alarm annunciation, they will go to the non-alarm state. Analog outputs and discrete control relays will be held at their current value or state when the operation starts.

At the completion of the AutoClean operation the hold will be released and normal operation will resume after the configured delay.

Procedure

If AutoClean was selected during I/O setup, follow the procedure in Table 5-8 to use the displayed screens for configuration when AUTO CLEAN is selected from the ADVANCED FEATURES configuration menu.

Table 5-8 Procedure for Configuring AutoClean

Step

1

2

Screen

MENU

CONFIGURATION

CALIBRATION

MAINTENANCE

I/O SETUP

DIAGNOSTICS

PREV NEXT

CONFIGURATION

CAT/RETRANSMISSION

ALARMS

DISCRETE CONTROL

ADVANCED FEATURES

DIAGNOSTICS

PREV NEXT

ADVANCED FEATURES

CLOCK SETUP

COMPUTED VALUE

AUTO CLEAN

DIAGNOSTICS

PREV NEXT

ATTENTION

24 HOUR CLOCK MUST

BE SETUP FOR THIS FEATURE

DIAGNOSTICS

NEXT

 

 

Action

Go to CONFIGURATION ADVANCED FEATURES

AUTO CLEAN.

If you have already set the clock, go to the next screen.

If you have not set the clock, press MENU to exit this function, then set the clock as described in Sub-Section 5.7.