Unpacking, Preparation, and Mounting
Honeywell
York, PA U.S.A.
ATTENTION - UNIT SET FOR 240 VAC
S/N: 9751Y712345670001
USA WARRANTY ASSISTANCE - 1 -
120/240 VAC 50/60 Hz 15VA
Contact Rating 1 & 2: 0.6 Amps at 120/240 VAC, 0.6 Amps at 110 VDC, and 2 Amps at 30 VDC.
If installed 3 & 4: 3 Amps at 120 VAC, 28 VDC. CE
CASE: NEMA 4X | Directions: |
Figure 3-1 Sample Nameplate
3.3 Mounting
Introduction
The 9782 Analyzer/Controller can be
Panel-mounting
Figures 3-2-1 and 3-2-2 (non-backlit) and 3-3-1 and 3-3-2 (backlit) show panel cutout dimensions. They also illustrate how the instrument is attached to a panel by clamping the edges of the cutout between the case flange and the supplied U-bracket which is fastened to the rear of the case.
The panel may be up to 3/ 8 in. thick. Cutouts for adjacent 9782 Analyzer/Controllers may be no closer than .687 in. horizontally and .937 in. vertically, resulting in a nominal edge-to-edge bezel separation of .078 in.
Note that if the panel already has a cutout for a 7082 instrument, the same cutout can be used for the 9782. If the panel has a cutout for a 7070, 7075, 7076, 7077, or 7078 instrument, a Reducer Panel Kit is available for mounting the 9782. The part number is in Section 11.
Wall- and pipe-mounting
A bracket is available as an option for wall or pipe mounting the unit as shown in Figures
The mounting bracket part number is in Section 11.
ATTENTION
Pipe mounting is not recommended if the pipe is subject to severe vibration. Excessive vibration may affect system performance.
7/99 | 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual |