Unpacking, Preparation, and Mounting

Honeywell

York, PA U.S.A.

ATTENTION - UNIT SET FOR 240 VAC

9782C-S0-VC-E0010-00

S/N: 9751Y712345670001

USA WARRANTY ASSISTANCE - 1 - 800-423-9883

120/240 VAC 50/60 Hz 15VA

Contact Rating 1 & 2: 0.6 Amps at 120/240 VAC, 0.6 Amps at 110 VDC, and 2 Amps at 30 VDC.

If installed 3 & 4: 3 Amps at 120 VAC, 28 VDC. CE

CASE: NEMA 4X

Directions: 70-82-25-74

Figure 3-1 Sample Nameplate

3.3 Mounting

Introduction

The 9782 Analyzer/Controller can be panel-, wall- or pipe-mounted.

Panel-mounting

Figures 3-2-1 and 3-2-2 (non-backlit) and 3-3-1 and 3-3-2 (backlit) show panel cutout dimensions. They also illustrate how the instrument is attached to a panel by clamping the edges of the cutout between the case flange and the supplied U-bracket which is fastened to the rear of the case.

The panel may be up to 3/ 8 in. thick. Cutouts for adjacent 9782 Analyzer/Controllers may be no closer than .687 in. horizontally and .937 in. vertically, resulting in a nominal edge-to-edge bezel separation of .078 in.

Note that if the panel already has a cutout for a 7082 instrument, the same cutout can be used for the 9782. If the panel has a cutout for a 7070, 7075, 7076, 7077, or 7078 instrument, a Reducer Panel Kit is available for mounting the 9782. The part number is in Section 11.

Wall- and pipe-mounting

A bracket is available as an option for wall or pipe mounting the unit as shown in Figures 3-2-1 and 3-2-2 (non-backlit) and 3-3-1 and 3-3-2 (backlit).

The mounting bracket part number is in Section 11.

ATTENTION

Pipe mounting is not recommended if the pipe is subject to severe vibration. Excessive vibration may affect system performance.

7/99

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

3-3

Page 45
Image 45
Honeywell 9782 Series manual Mounting, Panel-mounting, Wall- and pipe-mounting