Conductivity/Resistivity Analyzer/Controller
5.4Configuring Current Adjusting Type (CAT) Control and/or Retransmission of Process Variables
Purpose
If the hardware supports analog outputs, during I/O setup you had an opportunity to specify which process variable or computed value will be associated with each output. An analog output can be used for scaled retransmission of the value, or for Current Adjusting Type control by means of an output signal that is directly proportional to the input signal or computed value. Before using analog outputs, you must configure the related parameter values.
For a discussion of tuning proportional control see Appendix A.
Procedure
The procedure for configuring CAT control and/or retransmission of process variables is shown in Table
Table
Process Variables
Step
1
2
Screen
MENU
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP
DIAGNOSTICS
PREV NEXT
CONFIGURATION
CAT/RETRANSMISSION
ALARMS
DISCRETE CONTROL ADVANCED FEATURES
DIAGNOSTICS
PREV NEXT
CAT/RETRANSMISSION
CAT
RETRANSMISSION
DIAGNOSTICS
PREV NEXT
Action
Go to CONFIGURATION CAT/RETRANSMISSION.
ATTENTION: Any value entered for one of these parameters will take effect as soon as it is entered.
If the Analyzer/Controller hardware only supports one analog output, the screen at the left will be displayed when CAT/RETRANSMISSION is selected from the menu.5
If the hardware supports three analog outputs, the screen in Step 6 will be displayed when CAT/RETRANSMISSION is selected from the menu.
If CAT control will be used, select it from the menu.
5If the system uses a DE option card, no analog outputs are supported by the hardware. The screen will show “DE ENABLED”, instead of displaying analog outputs.
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual | 7/99 |