Conductivity/Resistivity Analyzer/Controller

Example 1: To determine the check resistor value needed to simulate conductivity measurement of 10 μS, use cell constant 0.1 and perform the following calculation:

10 k ohms = (0.1) x (1,000,000)

10

Example 2: To determine the check resistor value needed to simulate resistivity measurement of

10 M ohms, use cell constant 0.01 and perform the following calculation:

100 K ohms = (0.01) x (10,000,000)

Concentration values

If a wide range model is used for concentration measurement, obtain the appropriate check resistance value from the table below.

Data for Concentration Range Measurements

Material/Weight % Concentration

Hydrochloric Acid (HCl) 0 1 4

Sulfuric Acid (H2SO4) 0 1 4

Sodium Chloride (NaCl) 0 1 4

Sodium Hydroxide (NaOH) 0 1 4

Simulation Resistance (ohms) @ 25º C

Cell Constant

10

25

50

 

 

 

 

 

242.5

485.0

 

68.9

137.7

 

 

 

215.5

538.7

1077.4

56.0

140.0

280.0

 

 

 

574.1

1435.1

2870.3

195.2

398.0

796.1

 

 

 

189.2

473.0

946.1

54.0

135.1

270.1

 

 

 

Appendix D - 2 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

7/99