Specifications and Model Number

Wide range models

Wide Display Ranges: 9782C-W0

Conductivity (microsiemens/cm)

TDS

Cell Constant Required (cm-1)

 

 

 

100-1,999

0-1,999 ppm

0.1

 

 

 

1000-19,990

0-19,990 ppm

1

 

 

 

Conductivity (millisiemens/cm)

 

 

 

 

 

10-199.9

0-199.9 ppt

10

 

 

 

10-500

25

 

 

 

20-1,000

50

 

 

 

Concentration

 

 

 

 

 

0-18.00% Hydrochloric Acid*

 

50**

 

 

 

0-19.99% Sulfuric Acid

 

50**

 

 

 

0-19.99% Sodium Hydroxide*

 

25** ro 50

 

 

 

0-19.99% Sodium Chloride

 

25** or 50

 

 

 

*The maximum concentration measurable by conductivity depends on temperature. Maximum for sodium hydroxide allowed by the 9782 Analyzer is 10% at 0º C, 13% at 25º C and 20% at 75º C with linear interpolation between these points. The maximum concentration for hydrochloric acid is 15.5% above 40º C.

**Measurements in the lower half of the concentration ranges and below 50º C may also use the next lower available cell constant.

2.1.6 Operating Conditions

Operating Conditions

Ambient Temperature: Normal 0 to 60 °C (32 to 140 °F), extreme -10 to +60 °C (14 to 140 °F); storage -30 to +70°C (-22 to 158 °F).

Line Voltage: Normal 120 +/- 10% Vac (extreme 100 to 132 Vac, 47 to 63 Hz)

Normal 240 +/- 10% Vac (extreme 200 to 264 Vac, 50 to 60 Hz)

RH: 90% maximum non-condensing @ 40°C (104 °F) max.

Installation Category: II

Pollution Degree: 2

Altitude: <20,000 M (6562 ft)

7/99