Specifications and Model Number
Wide range models
Wide Display Ranges: 9782C-W0
Conductivity (microsiemens/cm) | TDS | Cell Constant Required |
|
|
|
0.1 | ||
|
|
|
1 | ||
|
|
|
Conductivity (millisiemens/cm) |
|
|
|
|
|
10 | ||
|
|
|
— | 25 | |
|
|
|
— | 50 | |
|
|
|
Concentration |
|
|
|
|
|
| 50** | |
|
|
|
| 50** | |
|
|
|
| 25** ro 50 | |
|
|
|
| 25** or 50 | |
|
|
|
*The maximum concentration measurable by conductivity depends on temperature. Maximum for sodium hydroxide allowed by the 9782 Analyzer is 10% at 0º C, 13% at 25º C and 20% at 75º C with linear interpolation between these points. The maximum concentration for hydrochloric acid is 15.5% above 40º C.
**Measurements in the lower half of the concentration ranges and below 50º C may also use the next lower available cell constant.
2.1.6 Operating Conditions
Operating Conditions
Ambient Temperature: Normal 0 to 60 °C (32 to 140 °F), extreme
Line Voltage: Normal 120 +/- 10% Vac (extreme 100 to 132 Vac, 47 to 63 Hz)
Normal 240 +/- 10% Vac (extreme 200 to 264 Vac, 50 to 60 Hz)
RH: 90% maximum
Installation Category: II
Pollution Degree: 2
Altitude: <20,000 M (6562 ft)
7/99 | 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual |