I/O Setup and System Configuration7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 5-215.9 Selecting a Computed Value to Be Displayed

Purpose

The Analyzer/Controller can calculate and display a computed value. The available values are:

RATIO: ratio between the values measured by the cells;

Cell
Cell
1
2

DIFFERENCE: difference between the values measured by the cells; Cell Cell12

%PASSAGE: computed percent passage:

Cell
Cell x
1
2100

%REJECTION: computed percent rejection: ()x1

1
2
100Cell
Cell

PPMCO2: computed parts per million carbon dioxide (standard range analyzers only)

Appendix B includes examples of applications using these values.

ATTENTION

If a computed value is used, it is significant which cell is considered “Cell 1” and which is

“Cell 2”. This must be taken into account when wiring the inputs. See 6.3 for details.

Procedure

Use the procedure in Table 5-9 to specify a computed value.

Table 5-9 Procedure for Specifying a Computed Value

Step Screen Action

1

MENU
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP
DIAGNOSTICS
| PREV | NEXT

Go to CONFIGURATION | ADVANCED FEATURES |

COMPUTED VALUE.

CONFIGURATION
CAT/RETRANSMISSION
ALARMS
DISCRETE CONT ROL
ADVANCED FE AT URES
DIAGNOSTICS
| PREV | NEXT
ADVANCED FEATURES
CLOCK SETUP
COMPUTED VA LU E
AUTO CLEAN
DIAGNOSTICS
| PREV | NEXT