I/O Setup and System Configuration

5.9 Selecting a Computed Value to Be Displayed

Purpose

The Analyzer/Controller can calculate and display a computed value. The available values are:

Cell1

RATIO: ratio between the values measured by the cells; Cell2

DIFFERENCE: difference between the values measured by the cells; Cell1 Cell2

 

 

Cell1

x100

 

 

%PASSAGE: computed percent passage: Cell2

 

 

(1

Cell1

)x100

 

 

%REJECTION: computed percent rejection: Cell2

PPMCO2: computed parts per million carbon dioxide (standard range analyzers only) Appendix B includes examples of applications using these values.

ATTENTION

If a computed value is used, it is significant which cell is considered “Cell 1” and which is “Cell 2”. This must be taken into account when wiring the inputs. See 6.3 for details.

Procedure

Use the procedure in Table 5-9 to specify a computed value.

Table 5-9 Procedure for Specifying a Computed Value

Step

1

Screen

MENU

CONFIGURATION

CALIBRATION

MAINTENANCE

I/O SETUP

DIAGNOSTICS

PREV NEXT

CONFIGURATION CAT/RETRANSMISSION ALARMS

DISCRETE CONTROL

ADVANCED FEATURES

DIAGNOSTICS

PREV NEXT

ADVANCED FEATURES CLOCK SETUP

COMPUTED VALUE

AUTO CLEAN

DIAGNOSTICS

PREV NEXT

Action

Go to CONFIGURATION ADVANCED FEATURES

COMPUTED VALUE.

7/99

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

5-21