I/O Setup and System Configuration
5.9 Selecting a Computed Value to Be Displayed
Purpose
The Analyzer/Controller can calculate and display a computed value. The available values are:
Cell1
∙RATIO: ratio between the values measured by the cells; Cell2
∙DIFFERENCE: difference between the values measured by the cells; Cell1 − Cell2
|
| Cell1 | x100 | ||
∙ |
|
| |||
%PASSAGE: computed percent passage: Cell2 | |||||
|
| (1 − | Cell1 | )x100 | |
∙ |
|
| |||
%REJECTION: computed percent rejection: Cell2 |
∙PPMCO2: computed parts per million carbon dioxide (standard range analyzers only) Appendix B includes examples of applications using these values.
ATTENTION
If a computed value is used, it is significant which cell is considered “Cell 1” and which is “Cell 2”. This must be taken into account when wiring the inputs. See 6.3 for details.
Procedure
Use the procedure in Table
Table 5-9 Procedure for Specifying a Computed Value
Step
1
Screen
MENU
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP
DIAGNOSTICS
PREV NEXT
CONFIGURATION CAT/RETRANSMISSION ALARMS
DISCRETE CONTROL
ADVANCED FEATURES
DIAGNOSTICS
PREV NEXT
ADVANCED FEATURES CLOCK SETUP
COMPUTED VALUE
AUTO CLEAN
DIAGNOSTICS
PREV NEXT
Action
Go to CONFIGURATION ADVANCED FEATURES
COMPUTED VALUE.
7/99 | 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual |