Conductivity/Resistivity Analyzer/Controller

Step

Screen

 

 

3AUTO CLEAN

TIMER SETUP

 

 

DAYS

3

 

HOURS

16

 

MINUTES

30

 

CURRENT DAY

2

 

DIAGNOSTICS

NEXT

 

4

 

 

 

 

 

AUTO CLEAN

 

 

 

 

RINSE SETUP

 

 

 

 

DURATION

180

 

 

 

RESUME DELAY

60

 

 

 

 

 

 

 

 

DIAGNOSTICS

 

 

 

 

PREV

NEXT

 

 

 

 

 

 

Action

Timer setup is used to specify the frequency of the rinses.

If several rinses are to be performed in a single day:

a)specify the number of days as zero

b)specify the number of hours and minutes in the interval between rinses.

If rinses are to occur at intervals of days:

a)specify the number of days in the interval

b)specify the time of day (hours and minutes) the

rinse should occur.

In our example, the rinse will occur every third day at 4:30 PM.

If you plan to use AutoClean, but initiate rinses manually using the real time AutoClean Time Remaining display:

a)specify the number of days as zero

b)specify the number of hours and minutes in the interval between rinses as zero also.

IF YOU ENTER ALL ZEROS, NO AUTOMATIC RINSES WILL BE PERFORMED.

Specify where today falls in the AutoClean cycle. For example, if the interval for rinses is 3 days and you want the first rinse to be tomorrow, specify that today is day 2.

Enter the duration of the rinse (in seconds).

Enter the number of seconds the Analyzer/Controller should wait after the rinse before resuming normal operation.

5-20

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

7/99

Page 76
Image 76
Honeywell 9782 Series manual Resume Delay Diagnostics Prev Next