Conductivity/Resistivity Analyzer/Controller
Step | Screen |
|
|
3AUTO CLEAN
TIMER SETUP |
|
|
DAYS | 3 |
|
HOURS | 16 |
|
MINUTES | 30 |
|
CURRENT DAY | 2 |
|
DIAGNOSTICS | ||
NEXT |
|
4 |
|
|
|
|
| AUTO CLEAN |
|
| |
|
| RINSE SETUP |
|
|
|
| DURATION | 180 |
|
|
| RESUME DELAY | 60 |
|
|
|
|
|
|
|
| DIAGNOSTICS |
|
|
|
| PREV | NEXT |
|
|
|
|
|
|
Action
Timer setup is used to specify the frequency of the rinses.
∙If several rinses are to be performed in a single day:
a)specify the number of days as zero
b)specify the number of hours and minutes in the interval between rinses.
∙If rinses are to occur at intervals of days:
a)specify the number of days in the interval
b)specify the time of day (hours and minutes) the
rinse should occur.
In our example, the rinse will occur every third day at 4:30 PM.
∙If you plan to use AutoClean, but initiate rinses manually using the real time AutoClean Time Remaining display:
a)specify the number of days as zero
b)specify the number of hours and minutes in the interval between rinses as zero also.
IF YOU ENTER ALL ZEROS, NO AUTOMATIC RINSES WILL BE PERFORMED.
Specify where today falls in the AutoClean cycle. For example, if the interval for rinses is 3 days and you want the first rinse to be tomorrow, specify that today is day 2.
Enter the duration of the rinse (in seconds).
Enter the number of seconds the Analyzer/Controller should wait after the rinse before resuming normal operation.
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual | 7/99 |